FDA Recall Terminated

BD BBL Vancomycin Screen Agar Catalog number 222204 Vancomycin Screen Agar is used to test enterococci for resistance to vancomycin and to predict the synergistic activity of this antimicrobial with an aminoglycoside antimicrobial.

Recall: Z-2628-2014 · Initiated July 28, 2014

Recall

Recall Number
Z-2628-2014
Event Number
69062
Firm
Becton Dickinson & Co. 7 Loveton Cir BD Diagnostic Systems
FEI Number
1119779
Product Code
JWY
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
July 28, 2014
Posted
September 12, 2014
Terminated
January 7, 2016
Address
Sparks, MD, 21152-9212

Description

BD BBL Vancomycin Screen Agar Catalog number 222204 Vancomycin Screen Agar is used to test enterococci for resistance to vancomycin and to predict the synergistic activity of this antimicrobial with an aminoglycoside antimicrobial.

Reason

These lots of Agar may have exhibited breakthrough growth of vancomycin susceptible Enterococcus species control organisms. Agar that allows susceptible enterococci to grow appears as a falsely resistant culture. If a lab does not QC, the error would be identified as falsely resistant enterococci would have further AST workup. There could be a delay in the identification of false-resistance

Action

BD Diagnostic System sent an Urgent Product Recall letter dated July 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Our records indicate you may have been shipped the above-referenced lot numbers of product between March 5, 2014 and June 5, 2014. Please discontinue use of these lot numbers and discard any remaining packages. If further assistance is needed, please contact BD Technical Services Department at 1-800-638-8663. The acknowledgement form included with the follow up letter that is faxed back to BD will be reconciled with the customer list. BD will contact 10% of non-respondents to verify receipt of the customer letter. BD has asked distributors to send their customer list to BD for follow up.

Distribution

Worldwide Distribution: US (Nationwide) including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI and WV; and Internationally to the countries of Belgium, Canada, Colombia, India, Mexico, Singapore and Taiwan.

Quantity

28540