FDA Recall Open, Classified

Allura Xper FD20 Biplane Product Codes: (1)722013, (2) 722008; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Recall: Z-2627-2025 · Initiated September 3, 2025

Recall

Recall Number
Z-2627-2025
Event Number
97501
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands
FEI Number
3002807880
Product Code
OWB
Status
Open, Classified
Root Cause
Device Design
Initiated
September 3, 2025
Posted
September 25, 2025

Description

Allura Xper FD20 Biplane Product Codes: (1)722013, (2) 722008; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Reason

Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

Action

On September 3, 2025, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. "Circulate this Urgent Medical Device Correction to all users of the system so that they are aware of the issue and follow the instructions included in the letter. 2. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction to that organization and inform Philips about this transfer through your local Philips representative. 3. Keep this Urgent Medical Device Correction with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. 4. Complete and return the response form included in this Urgent Medical Device Correction to Philips promptly and no later than 30 days from receipt. 5. If you experience the issue described in this letter, please report it to your local Philips representative. Actions Planned by Recalling Firm: 1. Philips will be replacing the Geo PC in all affected systems. 2. Philips expects to start replacing the Geo PC units by September 2025. Your local Philips representative will contact you to schedule a visit to confirm if your system has a model Wolfdale (Q35 generation) Geo PC and to plan the replacement of the Geo PC once available. 3. If you need additional information or support concerning these issues, contact your local Philips representative at 1-800-722-9377 (Philips Customer Care Solutions Center).

Distribution

Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Finland, France, FrenchPolynesia, Georgia, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.

Quantity

268 units (102 US, 166 OUS)