FDA Recall Terminated

ROTH MOBELI Grab Handles used with EOS System: MOBELI Stabi Vario Grab Handle, ROTH Catalogue Number 14002 96 S, EOS Imaging Catalogue Number 9000111773; and MOBELI Dual Grip Grab Handle, ROTH Catalogue Number 14002 25 S, EOS Imaging Catalogue Number 9000111774.

Recall: Z-2626-2017 · Initiated May 30, 2017

Recall

Recall Number
Z-2626-2017
Event Number
77424
FEI Number
3006165260
Product Code
IKX
Status
Terminated
Root Cause
Error in labeling
Initiated
May 30, 2017
Terminated
October 9, 2018
Address
EOS Imaging 4 Ieme Etage 10 rue Mercoeur Paris France

Description

ROTH MOBELI Grab Handles used with EOS System: MOBELI Stabi Vario Grab Handle, ROTH Catalogue Number 14002 96 S, EOS Imaging Catalogue Number 9000111773; and MOBELI Dual Grip Grab Handle, ROTH Catalogue Number 14002 25 S, EOS Imaging Catalogue Number 9000111774.

Reason

ROTH MOBELI grab handles could detach and patient could fall if the EOS system operator does not follow instructions. Before each use, the operator must examine the vacuum safety indicator to check the vacuum is monitored and sufficient and the grab handle correctly attached. Do not use grab handles to help patient up. The handles are only for stabilizing the patients hand or arm in the EOS cabin.

Action

EOS Imaging sent Field Safety Notice letter and Removal Instructions May 30, 2017. Customers are to immediately discontinue use of the grab handles and remove them from all affected units from inventory and return to a EOS Imaging. For further questions please call (678) 564-5400.

Distribution

US Distribution to the states of : Pennsylvania, Massachusetts, Minnesota, Alabama, Michigan, Texas and Washington.

Quantity

8 units each