FDA Recall Open, Classified

VITROS XT Chemistry Products ALKP Slides

Recall: Z-2624-2023 · Initiated August 5, 2023

Recall

Recall Number
Z-2624-2023
Event Number
92928
Firm
Ortho-Clinical Diagnostics, INc.
FEI Number
1000305840
Product Code
CJE
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
August 5, 2023
Posted
September 21, 2023
Address
513 Technology Blvd, Rochester, NY, 14626-3601

Description

VITROS XT Chemistry Products ALKP Slides

Reason

Samples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS ALKP Slide and not post the substrate deleption (DP) code, which may result in the System reporting an incorrect value (lower than the actual value), which may result in missed or delayed diagnosis of liver disease.

Action

An "IMPORTANT PRODUCT CORRECTION NOTIFICATION" letter dated 08/17/2023 was issued to impacted consignees. Consignees are provided instructions to utilize when reviewing sample results and any result between 600-1500 U/L should be diluted using a 10x dilution and retested. A confirmation of receipt of the recall letter should be completed and returned by August 25, 2023. Please forward this notification if the affected product was distributed outside your facility. Save this notification with your documentation or post this notification by each VITROS 250/350/FS 5,1/4600/5600/XT 3400/XT 7600 System until the issue has been resolved. If your laboratory has experienced this issue with the product and has not already done so, please report the occurrence to your local Ortho Care Technical Solutions Center. If you have further questions, please contact our Global Services Organization at 1-800-421-3311.

Distribution

Worldwide Distribution: US (nationwide) and OUS (foreign) distribution to countries of: Bermuda, Belgium, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, Italy, India, Japan, Mexico, Netherlands, Norway, Singapore, Spain, Sweden, Portugal, Russia, United Kingdom

Quantity

402,966 units