VITROS XT Chemistry Products TBIL-ALKP Slides
Recall
- Recall Number
- Z-2623-2023
- Event Number
- 92928
- Firm
- Ortho-Clinical Diagnostics, INc.
- FEI Number
- 1000305840
- Product Code
- CIG
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- August 5, 2023
- Posted
- September 21, 2023
- Address
- 513 Technology Blvd, Rochester, NY, 14626-3601
Description
VITROS XT Chemistry Products TBIL-ALKP Slides
Samples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS ALKP Slide and not post the substrate deleption (DP) code, which may result in the System reporting an incorrect value (lower than the actual value), which may result in missed or delayed diagnosis of liver disease.
An "IMPORTANT PRODUCT CORRECTION NOTIFICATION" letter dated 08/17/2023 was issued to impacted consignees. Consignees are provided instructions to utilize when reviewing sample results and any result between 600-1500 U/L should be diluted using a 10x dilution and retested. A confirmation of receipt of the recall letter should be completed and returned by August 25, 2023. Please forward this notification if the affected product was distributed outside your facility. Save this notification with your documentation or post this notification by each VITROS 250/350/FS 5,1/4600/5600/XT 3400/XT 7600 System until the issue has been resolved. If your laboratory has experienced this issue with the product and has not already done so, please report the occurrence to your local Ortho Care Technical Solutions Center. If you have further questions, please contact our Global Services Organization at 1-800-421-3311.
Worldwide Distribution: US (nationwide) and OUS (foreign) distribution to countries of: Bermuda, Belgium, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, Italy, India, Japan, Mexico, Netherlands, Norway, Singapore, Spain, Sweden, Portugal, Russia, United Kingdom
52,456 units