FDA Recall Open, Classified

BeneVision N1 Patient Monitor (N1). Part Numbers: 6660E-PA00017, 6660E-PA00014, 6660E-PA00012, 6660E-PA00011, 6660E-PA00018, 6660E-PA00015, 6660E-PA00020, Q6660E-PA00011, 6660E-PA00021, 6660E-PA00016, 6660E-PA00019, Q6660E-PA00018, 6660E-PA00010, Q6660E-PA00010, Q6660E-PA00021, Q6660E-PA00012, 6660E-PA00030, 6660E-PA00041, 6660E-PA00035, 6660E-PA00032, 6660E-PA00031. Software Versions: 1.04.00.01, 01.05.00.01, 01.06.00.01, and 01.06.01.01 The patient monitor is intended for monitoring, displaying, reviewing, storing, alarming and transferring of multiple physiological parameters.

Recall: Z-2622-2025 · Initiated August 15, 2025

Recall

Recall Number
Z-2622-2025
Event Number
97437
Firm
Mindray DS USA, Inc. dba Mindray North America
FEI Number
2221819
Product Code
MHX
Status
Open, Classified
Root Cause
Software design
Initiated
August 15, 2025
Posted
September 25, 2025
Address
800 Macarthur Blvd, Mahwah, NJ, 07430-2001

Description

BeneVision N1 Patient Monitor (N1). Part Numbers: 6660E-PA00017, 6660E-PA00014, 6660E-PA00012, 6660E-PA00011, 6660E-PA00018, 6660E-PA00015, 6660E-PA00020, Q6660E-PA00011, 6660E-PA00021, 6660E-PA00016, 6660E-PA00019, Q6660E-PA00018, 6660E-PA00010, Q6660E-PA00010, Q6660E-PA00021, Q6660E-PA00012, 6660E-PA00030, 6660E-PA00041, 6660E-PA00035, 6660E-PA00032, 6660E-PA00031. Software Versions: 1.04.00.01, 01.05.00.01, 01.06.00.01, and 01.06.01.01 The patient monitor is intended for monitoring, displaying, reviewing, storing, alarming and transferring of multiple physiological parameters.

Reason

Potential for activation of an abnormal alarm pause.

Action

Mindray contacted consignees via letter on about 08/15/2025. consignees were instructed that they may continue to use the BeneVision N1 patient monitor as a multi-parameter module with Mindray host monitors. Per the additional guidance, it is recommended to confirm that the remaining alarm time counter is actively displayed and follows the existing N1 Operator's Manual: Instructions for when the Alarm Pause is Activated. An image was also included in the notification letter showing the correctly displayed Alarm Pause notification and timer. Consignees were also requested to make all applicable clinical staff aware of the issue and to complete and return the Acknowledgement and Receipt form. Mindray's Care team will deploy a software update to correct the issue. Distributors were advised to contact their direct customers and notify them directly.

Distribution

Worldwide - Nationwide distribution and the country of Canada.

Quantity

2,278 units