Access Immunoassay Systems CORTISOL, Part Number: 33600
Recall
- Recall Number
- Z-2621-2010
- Event Number
- 55188
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- CGR
- Status
- Terminated
- Root Cause
- Other
- Initiated
- February 17, 2010
- Posted
- September 29, 2010
- Terminated
- April 19, 2012
- Address
- 200 S Kraemer Blvd, Brea, CA, 92821-6208
Description
Access Immunoassay Systems CORTISOL, Part Number: 33600
The recall was initiated by Beckman Coulter after the firm confirmed customer reports that the affected lot of Access Cortisol reagent (REF 33600) identified above may produce suppressed patient results. Beckman Coulter also confirmed customer reports that the two lots of Access Progesterone reagent (REF 33550) identified above may produce suppressed patient results and an increased incidence of I
Two (2) Product Corrective Action (PCA) letters with attached Customer Response form were sent on the week of February 15, 2010 to the affected customers. The letters provide the customers with an explanation of the problem identified and instruct the customers to discontinue use of Access Progesterone and Access Cortisol Reagents, lot numbers referenced. For product replacements consignees were instructed : (1) In the United States, please contact Customer Service at 1-800-526-3821; (2) In Canada, please contact Customer Service at 1-800-463-7828; (3) Outside of the United States and Canada, contact your local Beckman Coulter Representative. If consignees need assistance or have any questions regarding this notification, they were instructed to contact Technical Support at 1-800-854-3633 in the United States and Canada. Outside the United States and Canada, contact your local Beckman Coulter Representative.
United States & Canada
7282 units