FDA Recall Terminated

FluoroSet Radiographic Tubal Assessment Set Cannula, Manipulator/ Injector, Uterine Used for instillation of contrast media into the uterine cavity for radiographic evaluation of the uterine cavity and for injection of appropriate contrast media into the fallopian tubes for evaluation of tubal patency.

Recall: Z-2615-2016 · Initiated April 15, 2016

Recall

Recall Number
Z-2615-2016
Event Number
74010
Firm
Cook Inc.
FEI Number
1820334
Product Code
LFK
Status
Terminated
Root Cause
Unknown/Undetermined by firm
Initiated
April 15, 2016
Posted
August 22, 2016
Terminated
August 4, 2017
Address
750 N Daniels Way, Bloomington, IN, 47404-9120

Description

FluoroSet Radiographic Tubal Assessment Set Cannula, Manipulator/ Injector, Uterine Used for instillation of contrast media into the uterine cavity for radiographic evaluation of the uterine cavity and for injection of appropriate contrast media into the fallopian tubes for evaluation of tubal patency.

Reason

Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.

Action

Cook Inc. sent an URGENT: MEDICAL DEVICE RECALL NOTIFICATION letter dated April 15, 2016, to all affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with any medical questions or concerns, were instructed to contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. For information regarding the recall, please contact Stericycle Expert Solutions at 866.201.9067.

Distribution

Worldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, "Palestinian Territory, Occupied", Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, U.S. Virgin Islands, Yugoslavia.

Quantity

1,881