FDA Recall Open, Classified

Prucka 3 Amplifiers, Model P1801PA, and Field Replaceable Units (FRUs), Model 5875569, used with CardioLab AltiX and ComboLab AltiX.

Recall: Z-2613-2023 · Initiated August 8, 2023

Recall

Recall Number
Z-2613-2023
Event Number
92903
Firm
GE Medical Systems Information Technologies Inc
FEI Number
3012359877
Product Code
DQK
Status
Open, Classified
Root Cause
Component design/selection
Initiated
August 8, 2023
Posted
September 19, 2023
Address
9900 Innovation Dr, Wauwatosa, WI, 53226-4856

Description

Prucka 3 Amplifiers, Model P1801PA, and Field Replaceable Units (FRUs), Model 5875569, used with CardioLab AltiX and ComboLab AltiX.

Reason

A diode on the power supply of the Prucka 3 Amplifier used with CardioLab/ComboLab systems could reach elevated temperatures and fail, then power off and become inoperable.

Action

The recalling firm issued letters dated 8/8/2023 on the same day via FedEx. The letters explained the issue and listed actions to be taken by the customer/user. These included: (1) Ensure the facility has stable power. Lower voltage of the supplied power (e.g. 100V) causes more heat buildup and can cause the amplifier to fail; (2) Ensure the room temperature where the device is used is kept at < 30 degrees Celsius/86 degrees Fahrenheit. High temperatures (equal to or greater than 30 degrees Celsius/86 degrees Fahrenheit) can cause the amplifier to fail; (3) Ensure additional devices are available to monitor and/or stabilize the patient and/or complete the study (e.g., patient monitor, defibrillator, 3D Mapping system, etc.); and (4) Ensure that the hospital staff is familiar with utilizing the Direct Stimulator Connections on the CIM Block(s). This allows pacing from the stimulator when the amplifier has no power. The letter asks that all potential users in the consignee facility are made aware of this safety notification and recommended actions. GE Healthcare will correct all affected products and they will contact the consignee to plan for the hardware correction. An acknowledgment form was enclosed for completion and return no later than 30 days from receipt of the letter. The form is to acknowledge receipt and understanding of the letter and that the consignee has informed the appropriate staff and has taken appropriate actions. The form is to be completed by scanning or taking a photo of the completed form and emailing it back to the firm. A QR code is provided.

Distribution

Distribution was made to CA, CO, FL, GA, IL, IN, KY, MA, NC, NJ, OH, OR, PA, SC, TX, VA, and WI. There was government distribution but no military distribution. Foreign distribution was made to Canada, France, Germany, Greece, Ireland, Japan, Republic of Korea, and Romania.

Quantity

81 amplifiers