FDA Recall Terminated

Pressure Monitoring Kit, Catalog 16066100, Terumo Cardiovascular Systems Ann Arbor MI. The pressure monitoring kit is indicated for sterile single use with the reusable pressure transducer.

Recall: Z-2613-2011 · Initiated May 3, 2011

Recall

Recall Number
Z-2613-2011
Event Number
58838
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
1828100
Product Code
DTQ
Status
Terminated
Root Cause
Other
Initiated
May 3, 2011
Posted
June 22, 2011
Terminated
July 18, 2012
Address
6200 Jackson Road, Ann Arbor, MI, 48103-9586

Description

Pressure Monitoring Kit, Catalog 16066100, Terumo Cardiovascular Systems Ann Arbor MI. The pressure monitoring kit is indicated for sterile single use with the reusable pressure transducer.

Reason

The firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the pressure dome's manufacturing process.

Action

Terumo Cardiovascular Systems sent an URGENT MEDICAL DEVICE REMOVAL notice dated May 3, 2011, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. Customers were sent 10 replacement pressure monitoring kits with the removal notice as immediate replacements. The firm will send additional replacements to customers that have more than 10 units of affected product as noted on the Customers' returned reponse form. Customers that received the custom packs were provided stickers to identify the recalled packs. The customers were instructed not to open the packs to remove the affected pressure monitoring kit prior to clinical use, but to replace the pressure domes at the time of use. For more information please contact Terumo Cardiovascular Customer Service at 800- 521-2818.

Distribution

Worldwide Distribution -- USA (nationwide )including the states AR, CA, FL, GA, MO, MT, NC, ND, OK, OR, PA, TN, TX, UT, VA, WA and WY, and the countries of Argentina, Belgium, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Guatamala, Honduras, Hong Kong, Indonesia, Kuwait, Malaysia, Mexico, Pakistan, Panama, Philippines, Singapore, South Korea, Thailand, United Arab Emirates, Uruguay and Vietnam

Quantity

19495 units