FDA Recall Terminated

Dimension Vista Ferritin (FERR) Flex reagent cartridge

Recall: Z-2609-2017 · Initiated March 21, 2017

Recall

Recall Number
Z-2609-2017
Event Number
76915
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
DBF
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
March 21, 2017
Posted
May 5, 2017
Terminated
August 3, 2018
Address
500 Gbc Dr PO BOX 6101, Ms 514, Newark, DE, 19702-2466

Description

Dimension Vista Ferritin (FERR) Flex reagent cartridge

Reason

The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.

Action

Siemens mailed an Urgent Medical Device Correction Letter to affected customers on 03/29/2017 to inform them of the issue. The notice explained the reason for the field action, Risk to Health and actions to be taken by the customer.

Distribution

Product was distributed throughout the United States and Canada, United Arab Emirates, Argentina, Austria,Australia,Bangladesh,Belgium,Brazil,Canada Switzerland,China,Columbia,Czech Republic,Germany,Denmark,Algeria Egypt,Spain,Finland,France,Great Britain,Greece,Croatia,Indonesia,Ireland India,Italy,Japan,Canary Islands,Republic of Korea,Kuwait,Lebanon,Latvia Macedonia,Netherlands,Norway,New Zealand,Philippines,Poland,Portugal Qatar,Saudi Arabia,Slovenia,Slovakia,Turkey,Taiwan,South Africa

Quantity

29,266 units