Dimension Vista Ferritin (FERR) Flex reagent cartridge
Recall
- Recall Number
- Z-2609-2017
- Event Number
- 76915
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2517506
- Product Code
- DBF
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- March 21, 2017
- Posted
- May 5, 2017
- Terminated
- August 3, 2018
- Address
- 500 Gbc Dr PO BOX 6101, Ms 514, Newark, DE, 19702-2466
Description
Dimension Vista Ferritin (FERR) Flex reagent cartridge
The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.
Siemens mailed an Urgent Medical Device Correction Letter to affected customers on 03/29/2017 to inform them of the issue. The notice explained the reason for the field action, Risk to Health and actions to be taken by the customer.
Product was distributed throughout the United States and Canada, United Arab Emirates, Argentina, Austria,Australia,Bangladesh,Belgium,Brazil,Canada Switzerland,China,Columbia,Czech Republic,Germany,Denmark,Algeria Egypt,Spain,Finland,France,Great Britain,Greece,Croatia,Indonesia,Ireland India,Italy,Japan,Canary Islands,Republic of Korea,Kuwait,Lebanon,Latvia Macedonia,Netherlands,Norway,New Zealand,Philippines,Poland,Portugal Qatar,Saudi Arabia,Slovenia,Slovakia,Turkey,Taiwan,South Africa
29,266 units