FDA Recall Terminated

IGFBP-3 Immunoradiometric Assay (IRMA) Kit, Part Number: DSL-6600 usage: The DSL6600 IGFBP-3 immunoradiometric assay (IRMA) kit provides materials for the quantitative measurement of IGFBP-3 in serum. This assay is intended for in vitro diagnostic use.

Recall: Z-2607-2010 · Initiated August 9, 2010

Recall

Recall Number
Z-2607-2010
Event Number
56485
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
CFL
Status
Terminated
Root Cause
Pending
Initiated
August 9, 2010
Posted
September 28, 2010
Terminated
August 13, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92821-6208

Description

IGFBP-3 Immunoradiometric Assay (IRMA) Kit, Part Number: DSL-6600 usage: The DSL6600 IGFBP-3 immunoradiometric assay (IRMA) kit provides materials for the quantitative measurement of IGFBP-3 in serum. This assay is intended for in vitro diagnostic use.

Reason

The recall was initiated after Beckman Coulter confirmed customer reports that IGFBP-3 measurements using IGFBP-3 IRMA kit (REF DSL-6600) lots identified above may result in IGFBP-3 levels that are not in agreement with simultaneously measured levels of IGF-1. The investigation confirmed that above mentioned lots of IGFBP-3 kit provide results elevated by 84% on average. Erroneously elevated IGF

Action

The recall communication was initiated with a Product Corrective Action (PCA) letter that was sent on August 09, 2010 to all customers who received the IGFBP-3 IRMA kits, lot numbers referenced. The letter will provide the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to: (1) Discontinue using all lots identified above and discard in accordance with local regulations. (2) To continue testing for IGFBP-3 levels, find an alternate method. (3) Immediately notify all parties of interest (e.g., laboratory staff, clinicians) of the possibility of erroneously elevated IGFBP-3 results for patient samples tested with these lots, and advise all parties that clinical decision making and patient safety may be impacted. (4) Review patient IGFBP-3 results for the lots listed above, and interpret them in the light of additional tests (e.g., IGF-1) and the patient's clinical status and history. Customers were also asked to share the recall information with their laboratory staff and retain this notification as part of their laboratory Quality System documentation. Complete and return the enclosed response form within 10 days. For any questions regarding this notification, customers were to contact the DSL Technical Support Center at 1-800-231-7970 option 4 in the United States, or contact your local Beckman Coulter Representative.

Distribution

Nationwide in US and Canada

Quantity

63