FDA Recall Terminated

Panacea Air Pro Mattress Control Unit, Model: Platinum 5000, US Model Number 5000. Use with alternating pressure air flotation mattress.

Recall: Z-2605-2016 · Initiated July 11, 2016

Recall

Recall Number
Z-2605-2016
Event Number
74737
Firm
American National Mfg Inc
FEI Number
3000204953
Product Code
FNM
Status
Terminated
Root Cause
Software Manufacturing/Software Deployment
Initiated
July 11, 2016
Terminated
January 17, 2017
Address
252 Mariah Cir, Corona, CA, 92879-1751

Description

Panacea Air Pro Mattress Control Unit, Model: Platinum 5000, US Model Number 5000. Use with alternating pressure air flotation mattress.

Reason

The Panacea Air Pro Mattress Control Unit is being recalled due to a firmware problem, which may result in the device not having the firmness range required.

Action

An Urgent Medical Device Recall letter, dated 7/11/16, was sent to customers to inform them of a recall involving the Panacea Air Pro Mattress Control Unit also labeled as Platinum 5000, US Model Number 5000, Serial numbers 9500051 through 9500300. The letter informs the customers that the recall has been identified due to firmware problem. The letter informs the customers of the problem identified and the actions to be taken. Customers were asked to immediately examine their inventory and quarantine the affected device. A shipping label would be provided for return and update of the affected devices. All customers that the device was further distributed to would need to be notified of the recall.

Distribution

Distributed to the US -- WI, OH, IL, TX, AZ, MD, IA, PA, CO, WA, ID, MN, LA, WY, and MS.

Quantity

31 units