XiO Radiation Treatment Planning System, XiO Release 4.50.00 and above

Recall: Z-2604-2011 · Initiated May 19, 2010

Recall

Recall Number
Z-2604-2011
Event Number
58270
Firm
Computerized Medical Systems Inc
FEI Number
1937649
Product Code
MUJ
Status
Terminated
Root Cause
Software design
Initiated
May 19, 2010
Posted
June 20, 2011
Terminated
April 2, 2014
Address
13723 Riverport Drive, Suite 100, Maryland Heights, MO, 63043

Description

XiO Radiation Treatment Planning System, XiO Release 4.50.00 and above

Reason

XiO Software: when a treatment machine is defined in Source File Maintenance (SFM), the MLC Model is set to Elekta, and beams using this treatment machine are added to a new plan in Teletherapy, XiO automatically adds an MLC to the beam. This is because the Elekta linac will always have an MLC in place when the plan is actually delivered.

Action

A Safety Notice, dated 5/20/10, was sent to Elekta / Computerized Medical Systems Inc clients in May 2010. The notice indicates the problem, the clinical impact. The notice states a future XiO Release will resolve the issue. A return postcard is included with the mailing to verify consignee received the notice. For sites not returning the postcard by 5/28/10, the recalling firm will contact them via phone or email to confirm receipt of the notice. The notice also informed the consignees they would be contacted when a solution was available.

Distribution

Nationwide distribution to medical facilities including Puerto Rico. Distribution was also made to two Military and Government consignees. Foreign distribution was made to Australia, Belgium, Brazil, Canada, Colombia, France, India, Ireland, Israel, Italy, Malaysia, Netherlands, Philippines, Poland, Spain, and United Kingdom.

Quantity

57