FDA Recall Terminated

Focal Radiation Treatment Planning System, Focal Release 4.1.1 through 4.40.00. Product Usage: for use in radiation treatment planning using generally accepted contouring methods. The Focal software allows users to access XiO and other RTP systems from a remote PC-based system and provides the user with the ability to edit and modify patient data from a Windows-based PC and transfer the information back to the RTP system.

Recall: Z-2599-2011 · Initiated February 6, 2009

Recall

Recall Number
Z-2599-2011
Event Number
58250
Firm
Computerized Medical Systems Inc
FEI Number
1937649
Product Code
MUJ
Status
Terminated
Root Cause
Software design
Initiated
February 6, 2009
Posted
June 20, 2011
Terminated
April 8, 2014
Address
13723 Riverport Drive, Suite 100, Maryland Heights, MO, 63043

Description

Focal Radiation Treatment Planning System, Focal Release 4.1.1 through 4.40.00. Product Usage: for use in radiation treatment planning using generally accepted contouring methods. The Focal software allows users to access XiO and other RTP systems from a remote PC-based system and provides the user with the ability to edit and modify patient data from a Windows-based PC and transfer the information back to the RTP system.

Reason

Focal Software: if individual MLC leaf position edits are made, the original MLC aperture is retained. A subsequent aperture edit should not affect the MLC leaf positions however it was found that instead the leaves will then snap (unintentionally) back to the original aperture. Also, if the patient is sent to XiO after the MLC is edited, an isocenter shift made in XiO will also cause the MLCs to

Action

Elekta CMS Software sent an "ADVISORY NOTICE" dated February 6, 2009 to its affected customers. The notice identified the product, the problem, and the actions to be taken. Customers were informed that the workaround method available is to (edit the MLC pattern using the aperture first and the individual MLC leaf edits second). The firm states that this issue will be resolved in Focal version 4.50.00 and above. A "User Notice" will be sent to all users still using the affected software notifying them of the updated software available to correct the problem. If you have any questions, please call 314-993-0003 or toll free 800-878-4267.

Distribution

Worldwide Distribution -- USA (nationwide) including the state of: Puerto Rico and countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Egypt, Estonia, France, Germany, Hungary, India, Israel, Italy, Japan, Malaysia, Mexico, Mongolia, Morocco, Netherlands, New Zealand, Nicaragua, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sudan, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, United Kingdom, Uruguay and Venezuela.

Quantity

760