FDA Recall Open, Classified

FlexoStretch Model Number: 3156057 - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient.

Recall: Z-2598-2021 · Initiated August 11, 2021

Recall

Recall Number
Z-2598-2021
Event Number
88548
Firm
Hill-Rom, Inc.
FEI Number
1824206
Product Code
FSA
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
August 11, 2021
Address
1069 State Road 46 E, Batesville, IN, 47006-7520

Description

FlexoStretch Model Number: 3156057 - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient.

Reason

Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide backwards and potentially may fall out of the Oct lift sheet back onto the surface it is being lifted from.

Action

Hillrom issued URGENT Medical Device Correction letter dated August 11, 2021 by third party vendor via UPS mail/track the delivery to consignees. Letter states reason for recall, health risk and action to take: Hillrom is initiating this Medical Device Correction to all customers and will be correcting devices with an updated to the leveller. Actions being taken by Customer: Please share this Medical Device Correction with all potential users in your organization. Please ensure inspection of the device prior to use as stated in the device Instruction for Use (IFU). Complete the attached response form, acknowledging receipt, indicate the number of devices and return to [email protected] within one month

Distribution

Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, DE, IL, ME, MI, MO, MS, NJ, OH, PA, VA, WA and the countries of Argentina, Australia, Austria, Bolivia, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Latvia, Lithuania, Mexico, Netherlands, Norway, Poland, Portugal, Serbia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Province of China, Thailand, Turkey, United Arab Emirates.

Quantity

129 units