FDA Recall Terminated

Peloris II Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly, 3149 VIC, Australia; catalog number 26.0005. This tissue processor prepares tissue samples for sectioning by transforming fixed samples into wax embedded samples. This is achieved by exposing the tissue samples to a sequence of reagents in the processing retorts.

Recall: Z-2597-2011 · Initiated April 11, 2011

Recall

Recall Number
Z-2597-2011
Event Number
58502
Firm
Leica Microsystems, Inc.
FEI Number
3000210133
Product Code
IEO
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
April 11, 2011
Posted
June 20, 2011
Terminated
June 20, 2011
Address
2345 Waukegan Rd, Bannockburn, IL, 60015-1515

Description

Peloris II Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly, 3149 VIC, Australia; catalog number 26.0005. This tissue processor prepares tissue samples for sectioning by transforming fixed samples into wax embedded samples. This is achieved by exposing the tissue samples to a sequence of reagents in the processing retorts.

Reason

The processor potentially has a faulty impellor which could lead to reagent carry-over from one processing step to another,

Action

Leica Microsystems sent a MEDICAL DEVICE CORRECTION letter dated April 11, 2011, to the affected customers via first class mail. The letter identified the product, the problem, and the action needed to be taken by the customer. The Customers were instructed to discontinue use of the device until their local Leica Microsystems representative contact them to arrange the replacement of the faulty impellor. The accounts were asked to complete the attached Acknowledgement Form, listing the serial numbers of the units at their facility and fax it back to Leica at 1-847-607-3147. Additional questions should be addressed to the firm's office of Regulatory Affairs and Quality Assurance.

Distribution

Worldwide Distribution -- USA (nationwide) and the countries of Australia, Canada, Germany Japan, Portugal and United Kingdom.

Quantity

22 units