FDA Recall Terminated

MAYFIELD INFINITY XR2 LOW PROFILE BASE UNIT, Part Number A2079A The MAYFIELD Infinity XR2 Base Unit is intended to be used to support a patient during diagnostic examination and/or surgical procedures where a rigid support between surgical table and headrest, or skull clamp is necessary, positional freedom is required and where X-ray imaging modalities will be used.

Recall: Z-2594-2018 · Initiated June 29, 2018

Recall

Recall Number
Z-2594-2018
Event Number
80421
Firm
Integra LifeSciences Corp.
FEI Number
3003418325
Product Code
GDJ
Status
Terminated
Root Cause
Device Design
Initiated
June 29, 2018
Terminated
February 26, 2020
Address
311 Enterprise Dr, Plainsboro, NJ, 08536-3344

Description

MAYFIELD INFINITY XR2 LOW PROFILE BASE UNIT, Part Number A2079A The MAYFIELD Infinity XR2 Base Unit is intended to be used to support a patient during diagnostic examination and/or surgical procedures where a rigid support between surgical table and headrest, or skull clamp is necessary, positional freedom is required and where X-ray imaging modalities will be used.

Reason

There is a potential for the knob assembly in the swivel adapter to fracture/break during use.

Action

The firm, Integra, sent an "Urgent Voluntary Medical Device Correction Field Safety Notice" dated 6/29/18 to customers. The notification letter instructed customers to perform the following: 1) Review your MAYFIELD XR2 instructions for use. Please pay particular attention to the CAUTIONS listed on page 5 of the accompanying instructions for use section as noted below. " Always make sure the base unit is properly secured to the operating table " The base unit must not be used if the device appears to be damaged or functioning incorrectly " Over-tightening the mechanisms adjustment screws may result in damage to the unit " Over extending or overloading the base unit may result in unintended movement, shortened product life and/or damage to the unit " Do not alter the construction of this device as it may result in serious patient injury " Do not steam sterilize; components may be damaged by high heat leading to device damage and reduced performance 2) Visit Integra website and review video highlighting instructions for use at the following link: https://www.integralife.com/mayfield-imaging-cranial-stabilization-system/category/cranial-stabilization-mayfield-imaging-cranial-stabilization-system 3) Complete and return the attached Field Safety Notice Acknowledgement Form promptly via email: [email protected] or Fax to: 1-609-750-4220. Distributors were also instructed to: "We also kindly ask you to contact the final customers who may have the affected products and provide them with this letter." On July 20, 2018, an additional Urgent Voluntary Medical Device Recall letter was distributed to customers. The letters ask customers to perform the following: 1. Allow your sales representative to check the current knob assembly on your MAYFIELD Infinity XR2 Radiolucent Base Unit (A2079), Standard; MAYFIELD Infinity XR2 Low Profile Base Unit (A2079A); MAYFIELD Infinity XR2 Radiolucent Base Unit, Extended (A2079E) and MAYFIELD Spine Table Adaptor (

Distribution

Worldwide Distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV; and to countries of:: Australia, Canada, China, Dominican Republic, Hong Kong, India, Japan, Korea, Malaysia, Mexico, Pakistan, Panama, Peru, Singapore, Sri Lanka, Taiwan, and Thailand.

Quantity

1 units