FDA Recall Open, Classified

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ABDOMINAL VASCULAR-LF DYNJ905291B DYNJ905291C ARTERIOGRAM DYNJ28082G CVICU HEART BAG DYKM2130A ENDOVASCULAR PACK DYNJ0678934J EP PACK DYNJ23456J GENERAL ENDO PACK-LF DYNJ0842873K HYBRID PACK DYNJ65925B DYNJ65925C NON VASCULAR PACK DYNJ68037 PACEMAKER DYNJ901681D PACEMAKER PACK DYNJ51853C DYNJ51853D DYNJ60606A PACER PACK DYNJ40952A RADIOLOGY VASCULAR ACCESS TRAY DYNJVB1303A SM- INTERVENTIONAL DRAPE PK-LF DYNJ0780143K VASCULAR ACCESS PACK-LF DYNJ0101291D VASCULAR DSC DYNJ69172 DYNJ69172B VASCULAR HYBRID DYNJ57543F DYNJ57543I DYNJ57543J VASCULAR PACK DYNJ61702A

Recall: Z-2593-2026 · Initiated February 27, 2026

Recall

Recall Number
Z-2593-2026
Event Number
99062
FEI Number
1417592
Product Code
OEZ
Status
Open, Classified
Root Cause
No Marketing Application
Initiated
February 27, 2026
Posted
June 25, 2026
Address
3 Lakes Dr, Medline Industries, LP, Northfield, IL, 60093-2753

Description

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ABDOMINAL VASCULAR-LF DYNJ905291B DYNJ905291C ARTERIOGRAM DYNJ28082G CVICU HEART BAG DYKM2130A ENDOVASCULAR PACK DYNJ0678934J EP PACK DYNJ23456J GENERAL ENDO PACK-LF DYNJ0842873K HYBRID PACK DYNJ65925B DYNJ65925C NON VASCULAR PACK DYNJ68037 PACEMAKER DYNJ901681D PACEMAKER PACK DYNJ51853C DYNJ51853D DYNJ60606A PACER PACK DYNJ40952A RADIOLOGY VASCULAR ACCESS TRAY DYNJVB1303A SM- INTERVENTIONAL DRAPE PK-LF DYNJ0780143K VASCULAR ACCESS PACK-LF DYNJ0101291D VASCULAR DSC DYNJ69172 DYNJ69172B VASCULAR HYBRID DYNJ57543F DYNJ57543I DYNJ57543J VASCULAR PACK DYNJ61702A

Reason

Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.

Action

This entry catalogues SKUs communicated through the same letter documented in RES 98601 which were inadvertently missed in the initial report to the FDA. Firm began notifying customers on February 27, 2026 via letters titled "URGENT MEDICAL DEVICE RECALL". Customers were instructed to immediately check stock for affected product and quarantine any affected units. Upon receipt of the customers' response form reporting the quantity of affected product on hand, the firm will provide over-labels to place on affected inventory. Labels contain instructions for staff to remove and discard the affected component prior to using the kit.

Distribution

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.

Quantity

113, 843 kits