AMX Navigate Mobile X-Ray System

Recall: Z-2590-2021 · Initiated June 24, 2021

Recall

Recall Number
Z-2590-2021
Event Number
88601
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
IZL
Status
Terminated
Root Cause
Process control
Initiated
June 24, 2021
Posted
September 29, 2021
Terminated
June 14, 2023
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

AMX Navigate Mobile X-Ray System

Reason

An unexpected column motion while attempting to park the column on the AMX Navigate mobile X-ray system.

Action

On June 24, 2021 the firm contacted the facilities that received the device via email. An AMX Navigate system was provided to 3 healthcare facilities for evaluation under an agreement with GE Healthcare. While at the healthcare facilities, these systems remained under the ownership of GE Healthcare. After one of the healthcare facilities alerted GE Healthcare to a potential issue with its system, GE Healthcare removed the system from the 3 healthcare facilities in order to perform a detailed engineering analysis as part of its root cause investigation. These 3 systems have not been returned to the healthcare facilities and remain in GE Healthcare s possession.

Distribution

Domestic Distribution Only

Quantity

3 units