FDA Recall Terminated

Arkon Anesthesia Delivery System with Arkon Software Version 2.70, Display Unit Assembly 650-1769-00 installed, and Model #: 99999. This system includes a ventilator. The Spacelabs Arkon Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device is intended for use only by a suitably qualified physician.

Recall: Z-2589-2018 · Initiated July 11, 2018

Recall

Recall Number
Z-2589-2018
Event Number
80462
Firm
Spacelabs Healthcare, Ltd. Unit B Foxholes Centre; John Tate Road Hertford United Kingdom
FEI Number
3002807992
Product Code
CBK
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 11, 2018
Posted
August 10, 2018
Terminated
March 20, 2024

Description

Arkon Anesthesia Delivery System with Arkon Software Version 2.70, Display Unit Assembly 650-1769-00 installed, and Model #: 99999. This system includes a ventilator. The Spacelabs Arkon Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device is intended for use only by a suitably qualified physician.

Reason

Arkon Anesthesia Delivery System may go into a failed state (mechanical ventilation ceases) while the machine is in use or while idle.

Action

The firm, Spacelabs Healthcare, sent "Field Corrective Action Display Unit Resets" letters dated 7/3/2018 and 7/9/2018 to its customers via priority service on 07/11/2018. The letters informed them to circulate this notification to all affected persons and to add copies to appropriate user manuals. The firm will contact customers to replace the display assembly.. Customers with questions are encouraged to call Technical Support at 1-425-363-5212.

Distribution

Worldwide Distribution: U.S (nationwide) to states of.: LA, NE, CO, WY, MT, NM, IN, ME, NY, IL, MA, TX, PA, PR; and countries (OUS) of: PERU BOLIVIA, PANAMA, MALAYSIA, POLAND, IRELAND, MOROCCO, MEXICO, CANADA, UNITED KINGDOM, FRANCE, and CHINA, ECUADOR.

Quantity

328