CADD High Volume Administration Set with FlowStop, 21-7381-24. Product Usage: The CADD High-Volume Administration set is designed for use with the CADD Prizm pump to allow fluid deliver from an IV bag.
Recall
- Recall Number
- Z-2589-2017
- Event Number
- 77284
- Firm
- Smiths Medical ASD Inc.
- FEI Number
- 3012307300
- Product Code
- FPA
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- May 15, 2017
- Terminated
- April 4, 2018
- Address
- 6000 Nathan Ln N, Minneapolis, MN, 55442-1690
Description
CADD High Volume Administration Set with FlowStop, 21-7381-24. Product Usage: The CADD High-Volume Administration set is designed for use with the CADD Prizm pump to allow fluid deliver from an IV bag.
One lot of CADD High Volume Administration sets, part number 21-7381-24, lot number 26X035, incorrectly contained the IFU insert for part number 21-7357-24.
Consignees were sent on 5/15/2017 a Smiths Medical "Urgent Medical device Recall Notice" dated May 15, 2017. The letter described the Affected Devices, Reason for Recall, Risk to Health, & Instructions to Customers. Advised consignees to complete and send the Urgent Medical Device Recall Response Form to [email protected] within 10 calendar days of receipt of this Notice.If you have distributed affected devices to your customers, immediately notify your customers of this Recall and retrieve all affected devices. For questions contact Stericycle via email at: [email protected].
US Nationwide in the states of PA, TN, NV, CA, MD, AR, CA, VA, OH, MN, NY, VT, NY.
7,035