FDA Recall Terminated

CADD High Volume Administration Set with FlowStop, 21-7381-24. Product Usage: The CADD High-Volume Administration set is designed for use with the CADD Prizm pump to allow fluid deliver from an IV bag.

Recall: Z-2589-2017 · Initiated May 15, 2017

Recall

Recall Number
Z-2589-2017
Event Number
77284
Firm
Smiths Medical ASD Inc.
FEI Number
3012307300
Product Code
FPA
Status
Terminated
Root Cause
Process control
Initiated
May 15, 2017
Terminated
April 4, 2018
Address
6000 Nathan Ln N, Minneapolis, MN, 55442-1690

Description

CADD High Volume Administration Set with FlowStop, 21-7381-24. Product Usage: The CADD High-Volume Administration set is designed for use with the CADD Prizm pump to allow fluid deliver from an IV bag.

Reason

One lot of CADD High Volume Administration sets, part number 21-7381-24, lot number 26X035, incorrectly contained the IFU insert for part number 21-7357-24.

Action

Consignees were sent on 5/15/2017 a Smiths Medical "Urgent Medical device Recall Notice" dated May 15, 2017. The letter described the Affected Devices, Reason for Recall, Risk to Health, & Instructions to Customers. Advised consignees to complete and send the Urgent Medical Device Recall Response Form to [email protected] within 10 calendar days of receipt of this Notice.If you have distributed affected devices to your customers, immediately notify your customers of this Recall and retrieve all affected devices. For questions contact Stericycle via email at: [email protected].

Distribution

US Nationwide in the states of PA, TN, NV, CA, MD, AR, CA, VA, OH, MN, NY, VT, NY.

Quantity

7,035