FDA Recall Terminated

CT Goldseal BrightSpeed 16 PWR TIO 2 YR

Recall: Z-2587-2018 · Initiated May 24, 2018

Recall

Recall Number
Z-2587-2018
Event Number
80484
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
JAK
Status
Terminated
Root Cause
Component change control
Initiated
May 24, 2018
Terminated
February 8, 2019
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

CT Goldseal BrightSpeed 16 PWR TIO 2 YR

Reason

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Action

GE Healthcare notified customers on about 05/24/2018 via URGENT MEDICAL DEVICE CORRECTION letter. The letter outlined the reason for recall and notified customers that a GE Healthcare representative would be contacting them to arrange for the correction of the devices. Any questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or a local Service Representative.

Distribution

Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.

Quantity

42 units total