FDA Recall Open, Classified

Heartstring III Proximal Seal System. Intravascular anastomosis occluder.

Recall: Z-2586-2025 · Initiated August 15, 2025

Recall

Recall Number
Z-2586-2025
Event Number
97450
Firm
Maquet Cardiovascular, LLC
FEI Number
2242352
Product Code
DXC
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
August 15, 2025
Posted
September 18, 2025
Address
45 Barbour Pond Dr, Wayne, NJ, 07470-2094

Description

Heartstring III Proximal Seal System. Intravascular anastomosis occluder.

Reason

Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.

Action

Consignees were sent an URGENT MEDICAL DEVICE CORRECTION notification dated August 2025. The notification provides instructions for consignees to carry out in the instance of failure of the Heartstring Seal to load, failure of the Heartsring III Seal to deploy into the aortotomy, and failure of the deployed Heartstring III Seal to provide adequate hemostasis. Consignees are asked to forward the recall notification to individuals within their organization who use the recalled devices and to organizations to which product was further distributed. Consignees are to return the completed response form to [email protected]. Questions about this recall can be directed to Getinge Customer Service at 1-888-880-2874.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.

Quantity

126 units (91 US, 35 OUS)