FDA Recall Terminated

Operating room table - Product Usage: intended for use during diagnostic examinations, radiologic procedures or surgical procedures to support and position a patient.

Recall: Z-2585-2021 · Initiated August 25, 2021

Recall

Recall Number
Z-2585-2021
Event Number
88661
Firm
Deerfield Imaging, Inc.
FEI Number
3010326005
Product Code
KXJ
Status
Terminated
Root Cause
Device Design
Initiated
August 25, 2021
Terminated
April 30, 2024
Address
5101 Shady Oak Rd S, Minnetonka, MN, 55343-4100

Description

Operating room table - Product Usage: intended for use during diagnostic examinations, radiologic procedures or surgical procedures to support and position a patient.

Reason

ORT200/300 rotational lock may fail to function such that it cannot change states between locked and unlocked, due to a mechanical hardware issue.

Action

A Recall notification letter titled, "URGENT FIELD CORRECTION NOTICE" was sent to consignees on 08/25/2021 via Fed-Ex. E-mail communications were also sent to the customers on 8/26/2021. The letter instructs the consignee to fill out the second page of the notice and return it to the firm. The letter states that the IMRIS Customer Service will contact the consignee to schedule implementation of this corrective action with the facility. The letter mentions that, "IMRIS recommends these preventive actions: Per the OR Table Operator Manuals, the table should be tested daily in all standard positions and inspected for anomalies including the fit of the covers, tabletops, headrest, pendant connection and rotation. During usage, verify the table position is locked by manually pushing the table to ensure it does not exhibit rotational movement. If the rotational lock is observed to not function properly, please contact IMRIS Customer Service immediately."

Distribution

Worldwide distribution - US Nationwide distribution FL, NH, TX, OH, CA, OR, UT, TN, NC and the countries of Canada, Australia, Sweden, Japan.

Quantity

10 units