Philips SureSigns VS3 Vital Signs Monitor Models: VS3 863069, 863070, 453564041251, 453564041261, VS3 863071, 863072, 863073, 863074, 453564041271, 453564041281, 453564041291, 453564041301 The SureSigns VS3 vital signs monitor and VM4, VM6 and VM8 patient monitors are for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSigns VS3 vital signs monitor and VM4, VM6 and VM8 patient monitors are for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility.
Recall
- Recall Number
- Z-2584-2011
- Event Number
- 59037
- Firm
- Philips Healthcare Inc.
- FEI Number
- 1218950
- Product Code
- DPS
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- June 10, 2011
- Posted
- June 17, 2011
- Terminated
- August 15, 2013
- Address
- 3000 Minuteman Road, Andover, MA, 01810
Description
Philips SureSigns VS3 Vital Signs Monitor Models: VS3 863069, 863070, 453564041251, 453564041261, VS3 863071, 863072, 863073, 863074, 453564041271, 453564041281, 453564041291, 453564041301 The SureSigns VS3 vital signs monitor and VM4, VM6 and VM8 patient monitors are for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSigns VS3 vital signs monitor and VM4, VM6 and VM8 patient monitors are for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility.
Speakers on the SureSigns Patient Monitors and Vital Signs, failure may be identified by the display of a Speaker Malfunc message in the alarm message area or an AUDIO FAILED message in the main screen of the device or by the absence of audible sound
Philips Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated May 2011 to all affected customers. The letter identifies the products, problem, and actions to be taken. The letter states that customers may continue to use the affected products while awaiting replacement of the speaker, provided that the user routinely checks for the display of the "Speaker Malfunct." INOP. Additionally, the letter states that if the customer experience this INOP or there is no sound from the SureSigns Monitor, the customer is to remove the device from use and contact their local Philips service representative. The letter explains that Philips is completing the design for a replacement speaker assembly, which is expected to be available within 8 weeks. When available, Philips will replace the affected speaker assemblies free of charge. Customers are instructed to contact their local Philips Customer Care Center at 1-800-722-9377 if further information or support is required.
Worldwide Distribution--USA (nationwide) and countries of ARAB JAMAHIRI, AUSTRIA, AUSTRALIA, BAHRAIN, BANGLADESH, BELGIUM, CANADA ,CHILE, COLOMBIA, CROATIA, CYPRUS, FRANCE, GERMANY, HUNGARY, INDIA, IRELAND, ITALY, KUWAIT, LEBANON, LIBYAN, MALAYSIA, MEXICO, NAMIBIA, NETHERLANDS, NEW ZEALAND, OMAN, PHILIPPINES, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA ,SOUTH KOREA, SPAIN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, UNITED ARAB EMIRATES, and UNITED KINGDOM.
5225 units