XiO Radiation Treatment Planning System. XiO Release 4.50.00 and above Used to create treatment plans for any cancer patient for who external beam radiation therapy of brachytherapy has been prescribed.

Recall: Z-2582-2011 · Initiated November 16, 2010

Recall

Recall Number
Z-2582-2011
Event Number
57807
Firm
Computerized Medical Systems Inc
FEI Number
1937649
Product Code
MUJ
Status
Terminated
Root Cause
Other
Initiated
November 16, 2010
Posted
June 17, 2011
Terminated
April 3, 2014
Address
13723 Riverport Drive, Suite 100, Maryland Heights, MO, 63043

Description

XiO Radiation Treatment Planning System. XiO Release 4.50.00 and above Used to create treatment plans for any cancer patient for who external beam radiation therapy of brachytherapy has been prescribed.

Reason

XiO: When a new electron beam is added to a plan, the calculation Algorithm is set to Pencil Beam, and an Aperture is added to the beam, the user can select Port-Port Properties and edit the Material Thickness values. If the algorithm is then changed to Monte Carlo, the calculation will proceed and the user is not warned the Material/Thickness values just entered are not used in the dose calculati

Action

Elekta CMS Software drafted a User Notice which explains the reason for correction, the clinical impact, and the workaround method. The User Notice was distributed to all affected sites in November 2010. The issue will be resolved in the XiO Release 4.70.00, expected to be available in July 2011.

Distribution

Worldwide Distribution - USA (nationwide, Washington, DC and Puerto Rico and the countries of Albania, Algeria, Australia, Austria, Bahamas, Belarus, Belgium, Brazil, Canada, Chile, China,Croatia, Cyrus, Czech Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Kuwait, Latvia, Libya, Lithuania, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Phillipines,Poland, Portugal, Romoania, Russia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Suriname, Switzerland, Taiwan, Thailand, Turkey, Turmenistan, Ukraine, United Kingdom, and Vietnam