FDA Recall Terminated

OPEN HEART TRAY 1/CS Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc. .

Recall: Z-2581-2015 · Initiated July 10, 2015

Recall

Recall Number
Z-2581-2015
Event Number
71721
Firm
Customed, Inc Calle Igualdad Final # 7
FEI Number
2648727
Product Code
OFA
Status
Terminated
Root Cause
Packaging
Initiated
July 10, 2015
Posted
September 3, 2015
Terminated
April 12, 2017
Address
Fajardo, PR, 00738

Description

OPEN HEART TRAY 1/CS Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc. .

Reason

Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.

Action

Customed issued an "URGENT VOLUNTARY MEDICAL DEVICE RECALL" notification letter dated July 10, 2015 to all affected customers. The letter identified the product, problem, and actions to be taken by the customer.

Distribution

Distributed Only in Puerto Rico.

Quantity

56 trays