Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Customer Procedure Kit (Angiography/Angioplasty kit), REF: K05-02166B K05T-00380C K05T-01674B K05T-01863 K05T-01941C K05T-01955 K05T-02138 K05T-02159 K05T-02272G K05T-02272H K05T-02515A K05T-02533 K05T-02728 K05T-02733C K05T-03153 K05T-03201 K09T-00318D K09T-07304G K09T-07627L K09T-07627M K09T-08747 K09T-08774F K09T-08952 K09T-09037J K09T-09358 K09T-09576J K09T-09760C K09T-09978A K09T-10485L K09T-10728H K09T-10728J K09T-11022C K09T-11022D K09T-11379F K09T-11430B K09T-11478A K09T-11738D K09T-11798 K09T-11875B K09T-12020B K09T-12020C K09T-12200 K09T-12538F K09T-12538G K09T-12883B K09T-12883C K09T-12942B K09T-12942C K09T-12942D K09T-12942F K09T-12970B K09T-13293 K09T-13293A K09T-13294 K09T-13294A K09T-13384B K09T-13384C K09T-13444B K09T-13459A K09T-13459B K09T-13681A K09T-13740 K09T-13927 K09T-13927A K09T-13949 K09T-13972 K09T-14025 K09T-14028B K09T-14035 K09T-14072 K09T-14218 K10T-01528G K10T-01528H K10T-03769F K10T-04233J K10T-04941B K10T-05595A K10T-05611B K10T-05680B K10T-06085 K10T-06085A K10T-06101 K10T-12092 K10T-12092A K12-05727F K12T-01737D K12T-01816A K12T-01996K K12T-02064 K12T-02285F K12T-02667A K12T-02703C K12T-02703F K12T-02911F K12T-02911G K12T-02940 K12T-03196 K12T-03312 K12T-03724F K12T-03933C K12T-03975A K12T-03979C K12T-04085 K12T-04270A K12T-04271B K12T-04348A K12T-04383G K12T-04707 K12T-04926G K12T-05302D K12T-05303D K12T-05303F K12T-05327D K12T-05432F K12T-05432G K12T-05627 K12T-05849C K12T-05849D K12T-05849F K12T-05850B K12T-05850C K12T-05903B K12T-05968D K12T-05974A K12T-05996M K12T-05996N K12T-06043 K12T-06313B K12T-06412A K12T-06431A K12T-06436 K12T-06694A K12T-06821 K12T-06826 K12T-06839A K12T-07986C K12T-08071C K12T-08071D K12T-08072D K12T-08072F K12T-08166D K12T-08703 K12T-08964F K12T-09013F K12T-09237A K12T-09292A K12T-09338D K12T-09456B K12T-09840B K12T-10464 K12T-10514A K12T-10515 K12T-10515A K12T-10516 K12T-10562 K12T-10583A K12T-10586A K12T-10649 K12T-10664A K12T-10881B K12T-11049 K12T-11151 K12T-11586 K12T-11626B K12T-11626C K12T-11680 K12T-11684 K12T-11826 K12T-11853P K12T-11857B K12T-11880 K12T-11880A K12T-11880B K12T-11901 K12T-11903 K12T-11910 K12T-11923 K12T-11923A K12T-11924A K12T-11925A K12T-11943B K12T-11943C K12T-11984 K12T-11985 K12T-11996 K12T-12017 K12T-12091 K12T-12138 K12T-12149 K12T-12153 K12T-12155 K12T-12155A K12T-12155B K12T-12157 K12T-12157A K12T-12157B K12T-12224 K12T-12224A K12T-12224B K12T-12230 K12T-12242 K12T-12242A K12T-12244 K12T-12285A K12T-12291 K12T-12291A K12T-12323 K12T-12331A K12T-12397 K12T-12403 K12T-12406 K12T-12406A K12T-12407 K12T-12419 K12T-12448 K12T-12448A K12T-12494 K12T-11210 K12T-12321 K20T-00002 K20T-00004A K20T-00012 K20T-00016A K20T-00019 K20T-00027 K20T-00031 K20T-00035 K12T-02045 K12T-02909 K12T-03993A K12T-06280 K12T-09329A K12T-09368A K12T-10157 K12T-12095 K05T-01920F K09T-10485M K09T-14028 K09T-14028A K10-04922C K10T-00554J K10T-00554L K10T-00861D K10T-01428A K10T-01785L K10T-02735D K10T-03122A K10T-03348K K10T-03627F K10T-03627G K10T-03627H K10T-03776F K10T-03776G K10T-03784 K10T-03810A K10T-03810B K10T-03979M K10T-04060G K10T-04113Q K10T-04113R K10T-04160R K10T-04163H K10T-04199C K10T-04200C K10T-04256H K10T-04338G K10T-04415K K10T-04463D K10T-04500B K10T-04578B K10T-04655B K10T-04718F K10T-04723B K10T-04744D K10T-04774D K10T-04775 K10T-04780 K10T-04806A K10T-04844N K10T-04861B K10T-04862 K10T-04892 K10T-04983B K10T-04986F K10T-05012G K10T-05012H K10T-05081F K10T-05082G K10T-05165A K10T-05197B K10T-05219 K10T-05272C K10T-05307B K10T-05308B K10T-05365A K10T-05366B K10T-05367 K10T-05397 K10T-05451N K10T-05451R K10T-05462 K10T-05493B K10T-05504C K10T-05518 K10T-05531 K10T-05541C K10T-05541D K10T-05592B K10T-05593C K10T-05612 K10T-05651B K10T-05655 K10T-05662 K10T-05665A K10T-05666 K10T-05666A K10T-05667F K10T-05667G K10T-05667H K10T-05670B K10T-05711A K10T-05711B K10T-05746A K10T-05747A K10T-05797 K10T-05812A K10T-05817 K10T-05819
Recall
- Recall Number
- Z-2579-2024
- Event Number
- 94797
- Firm
- Merit Medical Systems, Inc.
- FEI Number
- 1721504
- Product Code
- OEQ
- Status
- Open, Classified
- Root Cause
- No Marketing Application
- Initiated
- June 11, 2024
- Posted
- August 12, 2024
- Address
- 1600 W Merit Pkwy South, Jordan, UT, 84095-2416
Description
Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Customer Procedure Kit (Angiography/Angioplasty kit), REF: K05-02166B K05T-00380C K05T-01674B K05T-01863 K05T-01941C K05T-01955 K05T-02138 K05T-02159 K05T-02272G K05T-02272H K05T-02515A K05T-02533 K05T-02728 K05T-02733C K05T-03153 K05T-03201 K09T-00318D K09T-07304G K09T-07627L K09T-07627M K09T-08747 K09T-08774F K09T-08952 K09T-09037J K09T-09358 K09T-09576J K09T-09760C K09T-09978A K09T-10485L K09T-10728H K09T-10728J K09T-11022C K09T-11022D K09T-11379F K09T-11430B K09T-11478A K09T-11738D K09T-11798 K09T-11875B K09T-12020B K09T-12020C K09T-12200 K09T-12538F K09T-12538G K09T-12883B K09T-12883C K09T-12942B K09T-12942C K09T-12942D K09T-12942F K09T-12970B K09T-13293 K09T-13293A K09T-13294 K09T-13294A K09T-13384B K09T-13384C K09T-13444B K09T-13459A K09T-13459B K09T-13681A K09T-13740 K09T-13927 K09T-13927A K09T-13949 K09T-13972 K09T-14025 K09T-14028B K09T-14035 K09T-14072 K09T-14218 K10T-01528G K10T-01528H K10T-03769F K10T-04233J K10T-04941B K10T-05595A K10T-05611B K10T-05680B K10T-06085 K10T-06085A K10T-06101 K10T-12092 K10T-12092A K12-05727F K12T-01737D K12T-01816A K12T-01996K K12T-02064 K12T-02285F K12T-02667A K12T-02703C K12T-02703F K12T-02911F K12T-02911G K12T-02940 K12T-03196 K12T-03312 K12T-03724F K12T-03933C K12T-03975A K12T-03979C K12T-04085 K12T-04270A K12T-04271B K12T-04348A K12T-04383G K12T-04707 K12T-04926G K12T-05302D K12T-05303D K12T-05303F K12T-05327D K12T-05432F K12T-05432G K12T-05627 K12T-05849C K12T-05849D K12T-05849F K12T-05850B K12T-05850C K12T-05903B K12T-05968D K12T-05974A K12T-05996M K12T-05996N K12T-06043 K12T-06313B K12T-06412A K12T-06431A K12T-06436 K12T-06694A K12T-06821 K12T-06826 K12T-06839A K12T-07986C K12T-08071C K12T-08071D K12T-08072D K12T-08072F K12T-08166D K12T-08703 K12T-08964F K12T-09013F K12T-09237A K12T-09292A K12T-09338D K12T-09456B K12T-09840B K12T-10464 K12T-10514A K12T-10515 K12T-10515A K12T-10516 K12T-10562 K12T-10583A K12T-10586A K12T-10649 K12T-10664A K12T-10881B K12T-11049 K12T-11151 K12T-11586 K12T-11626B K12T-11626C K12T-11680 K12T-11684 K12T-11826 K12T-11853P K12T-11857B K12T-11880 K12T-11880A K12T-11880B K12T-11901 K12T-11903 K12T-11910 K12T-11923 K12T-11923A K12T-11924A K12T-11925A K12T-11943B K12T-11943C K12T-11984 K12T-11985 K12T-11996 K12T-12017 K12T-12091 K12T-12138 K12T-12149 K12T-12153 K12T-12155 K12T-12155A K12T-12155B K12T-12157 K12T-12157A K12T-12157B K12T-12224 K12T-12224A K12T-12224B K12T-12230 K12T-12242 K12T-12242A K12T-12244 K12T-12285A K12T-12291 K12T-12291A K12T-12323 K12T-12331A K12T-12397 K12T-12403 K12T-12406 K12T-12406A K12T-12407 K12T-12419 K12T-12448 K12T-12448A K12T-12494 K12T-11210 K12T-12321 K20T-00002 K20T-00004A K20T-00012 K20T-00016A K20T-00019 K20T-00027 K20T-00031 K20T-00035 K12T-02045 K12T-02909 K12T-03993A K12T-06280 K12T-09329A K12T-09368A K12T-10157 K12T-12095 K05T-01920F K09T-10485M K09T-14028 K09T-14028A K10-04922C K10T-00554J K10T-00554L K10T-00861D K10T-01428A K10T-01785L K10T-02735D K10T-03122A K10T-03348K K10T-03627F K10T-03627G K10T-03627H K10T-03776F K10T-03776G K10T-03784 K10T-03810A K10T-03810B K10T-03979M K10T-04060G K10T-04113Q K10T-04113R K10T-04160R K10T-04163H K10T-04199C K10T-04200C K10T-04256H K10T-04338G K10T-04415K K10T-04463D K10T-04500B K10T-04578B K10T-04655B K10T-04718F K10T-04723B K10T-04744D K10T-04774D K10T-04775 K10T-04780 K10T-04806A K10T-04844N K10T-04861B K10T-04862 K10T-04892 K10T-04983B K10T-04986F K10T-05012G K10T-05012H K10T-05081F K10T-05082G K10T-05165A K10T-05197B K10T-05219 K10T-05272C K10T-05307B K10T-05308B K10T-05365A K10T-05366B K10T-05367 K10T-05397 K10T-05451N K10T-05451R K10T-05462 K10T-05493B K10T-05504C K10T-05518 K10T-05531 K10T-05541C K10T-05541D K10T-05592B K10T-05593C K10T-05612 K10T-05651B K10T-05655 K10T-05662 K10T-05665A K10T-05666 K10T-05666A K10T-05667F K10T-05667G K10T-05667H K10T-05670B K10T-05711A K10T-05711B K10T-05746A K10T-05747A K10T-05797 K10T-05812A K10T-05817 K10T-05819
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
On 06/11/2024, the firm sent an "URGENT MEDICAL DEVICE RECALL NOTICE" Letter to customers informing them that specific Merit Medical convenience kits contain plastic syringes manufactured by Jiangsu Shenli Medical Production Co. Ltd. These syringes lack the appropriate/sufficient FDA market authorization. In addition, Merit Medical posted the Urgent Medical Device Recall letter and complete list of affected products/kits on their website at https://www.merit.com/products/documents/product-notices/. Customer are instructed to: 1. Immediately determine if any of the products identified in the attached Customer Response Form (CRF) are within their facility, quarantine them, and discontinue use and distribution. 2. Ensure that appropriate personnel within their organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure this notice is immediately shared with them. Additional distribution details may be required by health authorities. 4. Follow the attached instructions for placing the recall labels on the products containing Jiangsu Shenli syringes, which inform the users to not use the affected syringes. If additional recall labels are needed, contact Customer Service at (801) 208-4381. Product may be returned to inventory once properly over-labeled. 5.At the point of use, the Jiangsu Shenli syringes are to be discarded. Photographs of Jiangsu Shenli syringes are enclosed to assist in identification. 6.Post this URGENT MEDICAL DEVICE RECALL NOTICE and Jiangsu Shenli Syringe photographs on or near the affected products. 7.If replacement syringes are needed, contact Customer Service at (801) 208-4381 for recommended syringe suppliers. For questions contact Merit Sales Representative or Merit Customer Service at 801-208-4381, Hours: 6 am to 6 pm MST Mon.-Fri.
U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
1,753,625 kits