FDA Recall
Terminated
Discovery MI Digital Ready
Recall: Z-2579-2018
·
Initiated May 24, 2018
Recall
- Recall Number
- Z-2579-2018
- Event Number
- 80484
- Firm
- GE Healthcare, LLC
- FEI Number
- 2126677
- Product Code
- KPS
- Status
- Terminated
- Root Cause
- Component change control
- Initiated
- May 24, 2018
- Terminated
- February 8, 2019
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
Discovery MI Digital Ready
Reason
DIGITAL ReadyCT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Action
GE Healthcare notified customers on about 05/24/2018 via URGENT MEDICAL DEVICE CORRECTION letter. The letter outlined the reason for recall and notified customers that a GE Healthcare representative would be contacting them to arrange for the correction of the devices. Any questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or a local Service Representative.
Distribution
Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
Quantity
42 units total