XiO Radiation Treatment Planning System, XiO Release 4.0.0 through 4.40.00 Used to create treatment plans for any cancer patient for who external beam radiation therapy or brachytherapy has been prescribed.
Recall
- Recall Number
- Z-2577-2011
- Event Number
- 58210
- Firm
- Computerized Medical Systems Inc
- FEI Number
- 1937649
- Product Code
- MUJ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- April 8, 2008
- Posted
- June 16, 2011
- Terminated
- April 8, 2014
- Address
- 13723 Riverport Drive, Suite 100, Maryland Heights, MO, 63043
Description
XiO Radiation Treatment Planning System, XiO Release 4.0.0 through 4.40.00 Used to create treatment plans for any cancer patient for who external beam radiation therapy or brachytherapy has been prescribed.
XiO: When the patient is positioned with their feet toward the CT gantry, the coordinates exported to the LAP IsoMark laser positioning system are not correct.
The firm, CMS, Inc., sent a "XiO Customer Advisory" letter dated April 8, 2008 to all affected customers. The letter describes the product, the problem, and the actions to be taken ( workaround). The letter also states the problem will be resolved in a future release. No release number or date was given. If you have any questions, please call 408-380-8023.
Worldwide distribution: USA (nationwide) and countries of: China, France, Germany, Japan, Mongolia, Netherlands, Nicaragua, Poland, Russia, South Korea, Switzerland, and Turkey.