FDA Recall Terminated

FOCAL Sim Workstation, Radiation Treatment Planning System. Used for patient contouring and viewing of treatment plan outputs. Also able to perform CT simulations and forward their information to the RTP System for dose calculation.

Recall: Z-2574-2014 · Initiated August 18, 2014

Recall

Recall Number
Z-2574-2014
Event Number
68984
Firm
Elekta, Inc.
FEI Number
1037831
Product Code
MUJ
Status
Terminated
Root Cause
Software design
Initiated
August 18, 2014
Posted
September 3, 2014
Terminated
August 8, 2017
Address
400 Perimeter Center Ter Ne, Ste 50, Atlanta, GA, 30346-1227

Description

FOCAL Sim Workstation, Radiation Treatment Planning System. Used for patient contouring and viewing of treatment plan outputs. Also able to perform CT simulations and forward their information to the RTP System for dose calculation.

Reason

When DICOM is exporting an arc plan (VMAT, Dynamic Conformal of 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported plan will not match the approved treatment plan.

Action

Elekta, Inc. issued an Urgent Important Field Safety Notice letter (IFSN 382-01-MON-001) to all affected customers.. Confirmation of receipt is to be completed by October 31, 2014. The problem will be resolved in a patch to Focal 4.80. Development is underway with a release date of September 2014 with confirmation of installation of the patch for the fix no later than 6 months from release of the patch. An updated notice from the manufacturer is also being sent to affected customers (FCO: 382-01-MON-001 VID:2.0). For questions contact your local Elekta office.

Distribution

Worldwide Distribution - US (nationwide)in the states of AK, CA, FL, MI, MN, MO, NJ, NY, NC, OH, PA, TN, VA,WA and Puerto Rico, Algeria, Antigua & Barbuda, Australia, Austria, Bahrain, Belarus, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech republic, Ecuador, Egypt, Estonia, France, Germany, Greece, India, Iraq, Israel, Italy, Japan, Kazakhstan, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Pakistan Portugal, Romania, Russia, Saudi Arabia, Singapore, Solvenia, South Africa, South Korea, Spain, Switzerland, Thailand, Turkey, and United Kingdom

Quantity

554 units