QC (Cs-137) nanoDot D2DNS, Model Numbers: a) 18135-000; b) 18135-KIT; c) 18135-SET; radiation monitoring dosimeter used with the microSTAR readers
Recall
- Recall Number
- Z-2569-2023
- Event Number
- 92647
- FEI Number
- 3008426232
- Product Code
- IYE
- Status
- Completed
- Root Cause
- Under Investigation by firm
- Initiated
- July 12, 2023
- Posted
- September 12, 2023
- Address
- Landauer 2 Science Rd, Glenwood, IL, 60425-1531
Description
QC (Cs-137) nanoDot D2DNS, Model Numbers: a) 18135-000; b) 18135-KIT; c) 18135-SET; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
Landauer disseminated an URGENT - Medical Device Recall notice to its consignees on 07/12/2023 by US mail and email. The notice explained the problem with the devices, potential risk, and requested the following: Discontinue use of the affected products, destroy used products, and return unused products.
Worldwide
10741 units