FDA Recall Terminated

BF-XP60 OES Bronchofiberscope, Model No. BF-XP60

Recall: Z-2566-2021 · Initiated August 16, 2021

Recall

Recall Number
Z-2566-2021
Event Number
88562
Firm
Olympus Corporation of the Americas
FEI Number
2429304
Product Code
EOX
Status
Terminated
Root Cause
Reprocessing Controls
Initiated
August 16, 2021
Terminated
March 11, 2024
Address
3500 Corporate Pkwy PO Box 610, Center Valley, PA, 18034-0610

Description

BF-XP60 OES Bronchofiberscope, Model No. BF-XP60

Reason

The firm is issuing validated, new reprocessing instructions according to the FDA 2015 guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."

Action

On August __, 2021, the firm distributed Urgent Medical Device Corrective Action letters to customers. The applicable addenda to the reprocessing instructions were attached to the letters. Action steps to be taken by the end user: 1. Inspect your inventory for the referenced devices and identify any device with the model names specified above. Please check all areas of the hospital to determine if any of these devices remain in inventory. 2. Carefully read the content of this Medical Device Correction Action as well as the attached Addendum . The addendum provides updated information or instructions on the chemicals to be used for reprocessing, the sterilization conditions, rinsing after high-level disinfection, and recommendation on using sterilization. 3. Ensure all personnel are completely knowledgeable and thoroughly trained on the new reprocessing steps. 4. Additional copies of the new Operation Manuals can be obtained by accessing our OlympusConnect site and downloading a copy of the new Manuals. Please visit our OlympusConnect customer website: at https://www.OlympusConnect.com New users will need to register. Once registered select [Product Support] on the left navigation bar, then select [Instruction Manuals], locate the applicable product codes and select the [Download] option. 5. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number and provide your contact information as indicated in the portal. 6. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. For further information, contact the firm's Technical Assistance Center at 1-800-848-9024, option 1.

Distribution

Domestic distribution nationwide. Product also distributed globally.

Quantity

180