Monaco Radiation Treatment Planning Workstation. Monaco Release 2.03.00 and above. The Monaco RTP system is used to create treatment plans for any cancer patient for whom external beam intensity modulated radiation therapy (IMRT) has been prescribed.
Recall
- Recall Number
- Z-2561-2011
- Event Number
- 57760
- Firm
- Computerized Medical Systems Inc
- FEI Number
- 1937649
- Product Code
- MUJ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- January 13, 2011
- Posted
- June 14, 2011
- Terminated
- April 3, 2014
- Address
- 13723 Riverport Drive, Suite 100, Maryland Heights, MO, 63043
Description
Monaco Radiation Treatment Planning Workstation. Monaco Release 2.03.00 and above. The Monaco RTP system is used to create treatment plans for any cancer patient for whom external beam intensity modulated radiation therapy (IMRT) has been prescribed.
Monaco: For VMAT plans, the measured dose of radiation delivered is lower than the software calculated dose by about 5% in Monaco Release 2.03.00 and 2.04.00.
Computerized Medical Systems, Inc. sent a "USER NOTICE" on January 13, 2011, to all affected users. This notice identified the product, the problem, and the action needed to be taken by the customer. A return postcard is included for the customer to confirm receipt of the user notification. The firm intends for correction in the dose calulation engine to appear in the Monaco Release 3.00 in March 2011. Any further questions please call (403) 830-8023.
Worldwide Distribution-- USA (nationwide) and countries of Austria, Australia, Belarus, Canada, China, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, India, Israel, Italy, Japan, Mexico, Netherlands, Poland, Singapore, Slovenia, South Korea, Spain, Switzerland, Taiwan, Thailand and UK.