1) Vascular Dilator, Sterile EO, Rx Only, Catalog #: DIL-100-12; 2) Dilator, Sterile EO, Rx Only, Catalog #: DIL-600-11; 3) Tearaway Introducer, Sterile EO, Rx Only, Catalog #: a) INT-028-14; b) INT-028-24; c) INT-029-05; d) INT-106-01; e) INT-106-14; 4) Micro-Access Tearaway Introducer Kit, Sterile EO, Rx Only, Catalog #: KIT-010-14; 5) Mini Tear Away Introducer Kit, Sterile EO, Rx Only, Catalog #: KIT-027-04; 6) LI-11 Plus G, Sterile EO, Catalog #: KIT-042-11; 7) LI-12 Plus G, Sterile EO, Catalog #: KIT-042-12; 8) Micro-Access Tearaway Introducer Kit, Sterile EO, Catalog #: a) KIT-900-13; b) KIT-900-14; c) KIT-900-17; d) KIT-900-29; e) KIT-901-02; f) KIT-901-02; 9) "Micro-Access" Kit, Sterile EO, Catolog #: KIT-900-88 10) Mini Stick II Microintroducer Kit. Sterile EO, Rx Only, Catalog #: a) KIT-045-05; b) KIT-045-15;
Recall
- Recall Number
- Z-2559-2018
- Event Number
- 80021
- Firm
- Galt Medical Corporation
- FEI Number
- 3000718472
- Product Code
- DYB
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 2, 2018
- Terminated
- July 20, 2020
- Address
- 2220 Merritt Dr, Garland, TX, 75041-6137
Description
1) Vascular Dilator, Sterile EO, Rx Only, Catalog #: DIL-100-12; 2) Dilator, Sterile EO, Rx Only, Catalog #: DIL-600-11; 3) Tearaway Introducer, Sterile EO, Rx Only, Catalog #: a) INT-028-14; b) INT-028-24; c) INT-029-05; d) INT-106-01; e) INT-106-14; 4) Micro-Access Tearaway Introducer Kit, Sterile EO, Rx Only, Catalog #: KIT-010-14; 5) Mini Tear Away Introducer Kit, Sterile EO, Rx Only, Catalog #: KIT-027-04; 6) LI-11 Plus G, Sterile EO, Catalog #: KIT-042-11; 7) LI-12 Plus G, Sterile EO, Catalog #: KIT-042-12; 8) Micro-Access Tearaway Introducer Kit, Sterile EO, Catalog #: a) KIT-900-13; b) KIT-900-14; c) KIT-900-17; d) KIT-900-29; e) KIT-901-02; f) KIT-901-02; 9) "Micro-Access" Kit, Sterile EO, Catolog #: KIT-900-88 10) Mini Stick II Microintroducer Kit. Sterile EO, Rx Only, Catalog #: a) KIT-045-05; b) KIT-045-15;
The products listed might contain unsafe levels of bacterial endotoxins (Pyrogens) that were introduced during a manufacturing step. Bacterial Endotoxins also called pyrogenic bacteria can activate the inflammatory process and produce fever, chills, and hypotension in a patient.
On approximately 05/01/2018 recall letters and customer response forms were mailed to all consignees.
U.S.: MN, PA, VA, CA, UT, FL, NY, GA, LA, MO, TX, MS, OH, WA, MA, IL, KY, NC, and Puerto Rico; OUS: Canada, Australia, France, Germany, Ireland, Italy, Netherlands, Poland, Slovakia, Switzerland, Belgium, United Kingdom, Korea
9050