FDA Recall Terminated

Leica Microsystems products labeled as the following: Leica M320 Microscope Systems with Article Numbers 10448419, 10448420, 10448485, 10448486 Product Usage: Intended use of this device is an optical instrument for magnifying and illuminating specimens. It can be applied for observation and documentation and for human and veterinary medical treatment. Not for use in Eye Operations.

Recall: Z-2555-2018 · Initiated June 4, 2018

Recall

Recall Number
Z-2555-2018
Event Number
80229
Firm
Leica Microsystems, Inc.
FEI Number
3000210133
Product Code
EPT
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
June 4, 2018
Terminated
March 25, 2021
Address
1700 Leider Ln, Buffalo Grove, IL, 60089-6622

Description

Leica Microsystems products labeled as the following: Leica M320 Microscope Systems with Article Numbers 10448419, 10448420, 10448485, 10448486 Product Usage: Intended use of this device is an optical instrument for magnifying and illuminating specimens. It can be applied for observation and documentation and for human and veterinary medical treatment. Not for use in Eye Operations.

Reason

Possible failure of screws which hold the gas spring fixation bracket in place. Due to this failure, the gas spring, which is fixed to the upper part of the parallelogram, cannot hold up the weight of the optics carrier.

Action

Leica Microsystems notified customers on about 06/04/2018 via URGENT FIELD SAFETY NOTICE letter. The letter informed customers that they would be contacted by a Leica Microsystems representative to arrange for the field update in the next 12 months. Customers were instructed that they could continue to use the system pending the update if they ensure a functional check is performed prior to surgery and follow user manual directions. If any anomalies are noticed prior to the correction, customers should contact their Leica representative. Customers were also instructed to provide the notification to all appropriate personnel , notify customers if the unit was transferred, and complete and return the Medical Device Correction Form. A follow-up recall notification was sent to customers on or about 06/07/2018. The follow up letter discusses the functional check prior to use and provides additional compliant information.

Distribution

Worldwide Distribution US Nationwide in the states of : CA, CO, CT, FL, HI, ID, IL, KS, KY, LA, MD, MI, MN, MO, NC, NJ, NY, OH, PA, SD, TN, TX. International Distribution: Canada, AUSTRALIA, BELGIUM, Bolivia, BRAZIL, BULGARIA, CHILE, CHINA, DENMARK, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, HONGKONG, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KUWAIT, MALAYSIA, PHILIPPINES, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, TAIWAN,THAILAND, TURKEY, UK

Quantity

8,564 Total Units