CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor, MODEL # CSM-BC400; --- COMMON/USUAL NAME: Bed Patient Monitor; CLASSIFICATION NAME: Bed Patient Monitor --- Monitor is connected to a sensor pad that monitors the patient if they attempt to arise from a bed, chair, floor mat, or toilet seat. The monitor will sound an alarm.
Recall
- Recall Number
- Z-2554-2014
- Event Number
- 68961
- Firm
- Curbell Medical, Inc.
- FEI Number
- 3000210521
- Product Code
- KMI
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- August 5, 2014
- Posted
- August 26, 2014
- Terminated
- February 28, 2017
- Address
- 20 Centre Dr, Orchard Park, NY, 14127-4102
Description
CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor, MODEL # CSM-BC400; --- COMMON/USUAL NAME: Bed Patient Monitor; CLASSIFICATION NAME: Bed Patient Monitor --- Monitor is connected to a sensor pad that monitors the patient if they attempt to arise from a bed, chair, floor mat, or toilet seat. The monitor will sound an alarm.
The firm became aware of a potential problem that was initiated by a customer complaint. After consultation with the manufacturer, it was discovered that a resistor was incorrectly placed within the circuit board on the monitor. This change to the resistor was a planned change to address a product improvement (improve battery drain). However, this modification rendered the monitor's alarm via the nurse call communication port to not function.
Urgent Medical Device Recall Letters (dated 8/01/14) and Response Forms were sent to the consignees via e-mail (w/read receipt) on 8/05/14.
Nationwide, Canada, Ireland, Italy
Domestic: 112 units; Foreign: 211 units