FDA Recall Terminated

804115, Terumo Coronary Ostia cannula, 10 Fr, 10 (25 cm) long A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

Recall: Z-2554-2011 · Initiated May 16, 2011

Recall

Recall Number
Z-2554-2011
Event Number
58762
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
1828100
Product Code
DWF
Status
Terminated
Root Cause
Process control
Initiated
May 16, 2011
Posted
June 14, 2011
Terminated
March 21, 2012
Address
6200 Jackson Road, Ann Arbor, MI, 48103-9586

Description

804115, Terumo Coronary Ostia cannula, 10 Fr, 10 (25 cm) long A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

Reason

Terumo has discovered several instances of adhesive build-up and or skiving of plastic (plastic flash) inside the distal tip of the Terumo Coronary Ostia Cannula. The defect was discovered during in-house inspection after Terumo CVS implemented a new inspection process for the cannula. There have been no reports of the defect in clinical use.

Action

Terumo Cardiovascular Systems Corporation sent an URGENT MEDICAL DEVICE REMOVAL notice dated May 16, 2011, to all affected customers. The notice identified the product, the problem, and the action to be taken. The letter described the problem, the potiential hazard, and the action to be taken by the consignee. Consignees were instructed to discontinue use of the product and return all product to Terumo. The letter provided a list of alternate products to be used instead of the recalled products. Consignees with questions should call 1-800-521-2818 Monday through Friday 8AM to 6PM.

Distribution

Worldwide Distribution - USA (nationwide) and the countries of BELGIUM, BRAZIL, DOMINICAN REPUBLIC, UNITED ARAB EMIRATES (UAE), SINGAPORE, MALAYSIA, AND SOUTH AFRICA.

Quantity

900