Siemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices intended to measure a variety of analytes in human body fluids. All Serial #s of material #s: 10636929, 10636928, 10486890, 10472176, 10639337, and 10639338.
Recall
- Recall Number
- Z-2553-2014
- Event Number
- 68854
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2517506
- Product Code
- JJE
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- July 16, 2014
- Posted
- August 26, 2014
- Terminated
- December 16, 2014
- Address
- 500 Gbc Dr PO BOX 6101, Ms 514, Newark, DE, 19702-2466
Description
Siemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices intended to measure a variety of analytes in human body fluids. All Serial #s of material #s: 10636929, 10636928, 10486890, 10472176, 10639337, and 10639338.
The Reagent Lid hinge may lose its effectiveness and slowly shift downward during maintenance procedures, potentially resulting in physical contact and injury to the operator.
Siemens notified end users of a potential safety issue with the Reagent Lid hinge on the Dimension EXL instrument via an Urgent Medical Device Correction letter, dated July 2014. The letter indicated Customer Service Engineers will inspect and adjust the tension on the reagent lid hinges during every visit and replace hinges which no longer properly adjust. Users were urged to use caution when the Reagent Lid is open and notify Siemens the Reagent Lid is shifting down.
Nationwide and internationally, including Canada and Mexico
2757