FDA Recall Terminated

AirLife Resuscitation Device, Adult labeled as the following: a. Adult, with mask, oxygen reservoir bag, Part Number 2K8004; b. Adult, with mask, oxygen reservoir bag, CO2 detector, PEEP valve, Part Number 2K8035C2; c. Adult, with mask, 28" large bore tubing, Part Number 2K8017; d. Adult, with mask, 40" oxygen reservoir tubing, Part Number 2K8005; e. Adult, with mask, 40" oxygen reservoir tubing, PEEP valve, Part Number 2K8036; f. Adult, without mask, 40" oxygen reservoir tubing, Part Number 2K8001; g. Adult, with mask, oxygen reservoir bag, manometer, PEEP valve, Part Number 2K8035M; h. Adult, with mask, oxygen reservoir bag, CO2 detector, Part Number 2K8004C2; Manual resuscitation devices that are used to provide positive pressure ventilation to patients who are not breathing or cannot adequately breathe on their own. The devices are intended for use as single patient use resuscitator devices for pulmonary resuscitation.

Recall: Z-2552-2018 · Initiated May 8, 2018

Recall

Recall Number
Z-2552-2018
Event Number
80500
Firm
Vyaire Medical
FEI Number
2050001
Product Code
BTM
Status
Terminated
Root Cause
Equipment maintenance
Initiated
May 8, 2018
Posted
August 8, 2018
Terminated
July 7, 2020
Address
26125 N Riverwoods Blvd, Mettawa, IL, 60045-3420

Description

AirLife Resuscitation Device, Adult labeled as the following: a. Adult, with mask, oxygen reservoir bag, Part Number 2K8004; b. Adult, with mask, oxygen reservoir bag, CO2 detector, PEEP valve, Part Number 2K8035C2; c. Adult, with mask, 28" large bore tubing, Part Number 2K8017; d. Adult, with mask, 40" oxygen reservoir tubing, Part Number 2K8005; e. Adult, with mask, 40" oxygen reservoir tubing, PEEP valve, Part Number 2K8036; f. Adult, without mask, 40" oxygen reservoir tubing, Part Number 2K8001; g. Adult, with mask, oxygen reservoir bag, manometer, PEEP valve, Part Number 2K8035M; h. Adult, with mask, oxygen reservoir bag, CO2 detector, Part Number 2K8004C2; Manual resuscitation devices that are used to provide positive pressure ventilation to patients who are not breathing or cannot adequately breathe on their own. The devices are intended for use as single patient use resuscitator devices for pulmonary resuscitation.

Reason

An error in manufacturing of the AirLife Resuscitation Devices that has the potential to obstruct airflow of the supplemental oxygen delivery during use.

Action

Vyaire Medical notified customers on about 05/08/2018 via URGENT RECALL NOTIFICATION letter. Instructions included to conduct 100% physical inventory for any affected devices, destroy any affected devices according to each facility's destruction protocol, notify customers of the recall if affected devices have been further distributed, and complete and return the Customer Response Form. Questions and comments can be directed to Vyaire Medical's Customer Support Team at 1-833-327-3284 (Option #1) or email: [email protected] Monday-Friday 8:00 am - 5:30 pm CST.

Distribution

US distribution only to: AL, AR, AZ, CA, CA, CO, FL, GA, HI, IA, IL, KY, LA, MA, MD, MI, MN, MO, MS, NC,NJ, NY, OH, OR, PR, SD, TX, VA, WA.

Quantity

15,714 units total