FDA Recall Terminated

MOL3650 Simplexa VZV Direct - Product Usage: is intended for use on the LIAISON MDX instrument for the qualitative detection of varicella-zoster virus (VZV) DNA in cerebrospinal fluid (CSF) from patients with signs and symptoms of VZV infections of the central nervous system (CNS).

Recall: Z-2548-2020 · Initiated May 12, 2020

Recall

Recall Number
Z-2548-2020
Event Number
85744
Firm
DiaSorin Molecular LLC
FEI Number
2023365
Product Code
PLO
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
May 12, 2020
Posted
July 10, 2020
Terminated
April 19, 2024
Address
11331 Valley View St, Cypress, CA, 90630-5366

Description

MOL3650 Simplexa VZV Direct - Product Usage: is intended for use on the LIAISON MDX instrument for the qualitative detection of varicella-zoster virus (VZV) DNA in cerebrospinal fluid (CSF) from patients with signs and symptoms of VZV infections of the central nervous system (CNS).

Reason

Due to a software malfunction in the Multi Assay Suite, there is the probability of increased occurrence of false negative results.

Action

On 05/12/2020, the firm sent an "URGENT FIELD SAFETY NOTIFICATION LETTER" with an acknowledgement form to customers and distributors via email to inform them of an internal observed issue associated with the use of their Multi-Assay Suite (MAS) featured in the software version 2.0 where it may incorrectly apply the thresholds of the first loaded assay to the analytes coupled with the same dye in the second assay resulting in false negative or false positive results. The firm is instructing customers to: -Immediately discontinue the use of the MAS feature and that the issue will be correct in the new Software Version 2.1 which is expected to be release in June 2020. -Fill out the Response Form and return by either: -fax (562) 240-6526 Attention: Technical Services -email to: technical [email protected] -mail to: ATTN: Technical Services, 11331 Valley View Street, Cypress, CA 90630 -Forward this communication to all required individuals with their organization. For further assistance with evaluating assay runs or for any questions, to contact Technical Services at +1-5672-240-6500, option 3 or via email [email protected]

Distribution

Worldwide distribution - U.S. Nationwide distribution including in the states of CO, FL, KS, MI, NC, PA, TX, and WA The countries of Austria, Czech Republic, France and Switerland.

Quantity

405 plastic vials (24 vials per box)