FDA Recall Terminated

MOL2150 Simplexa HSV 1&2 Direct -CSF -Genital Swab -Cutaneous / Mucocutaneous Swabs Product Usage: is intended for use on the LIAISON MDX instrument for the qualitative detection and differentiation of HSV-1 and HSV-2 present in cerebrospinal fluid (CSF), or cutaneous and mucocutaneous lesion swabs from patients with signs and symptoms of HSV-1 or HSV-2 infection of the central nervous system (CNS) or cutaneous and mucocutaneous lesions.

Recall: Z-2547-2020 · Initiated May 12, 2020

Recall

Recall Number
Z-2547-2020
Event Number
85744
Firm
DiaSorin Molecular LLC
FEI Number
2023365
Product Code
OQO
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
May 12, 2020
Posted
July 10, 2020
Terminated
April 19, 2024
Address
11331 Valley View St, Cypress, CA, 90630-5366

Description

MOL2150 Simplexa HSV 1&2 Direct -CSF -Genital Swab -Cutaneous / Mucocutaneous Swabs Product Usage: is intended for use on the LIAISON MDX instrument for the qualitative detection and differentiation of HSV-1 and HSV-2 present in cerebrospinal fluid (CSF), or cutaneous and mucocutaneous lesion swabs from patients with signs and symptoms of HSV-1 or HSV-2 infection of the central nervous system (CNS) or cutaneous and mucocutaneous lesions.

Reason

Due to a software malfunction in the Multi Assay Suite, there is the probability of increased occurrence of false negative results.

Action

On 05/12/2020, the firm sent an "URGENT FIELD SAFETY NOTIFICATION LETTER" with an acknowledgement form to customers and distributors via email to inform them of an internal observed issue associated with the use of their Multi-Assay Suite (MAS) featured in the software version 2.0 where it may incorrectly apply the thresholds of the first loaded assay to the analytes coupled with the same dye in the second assay resulting in false negative or false positive results. The firm is instructing customers to: -Immediately discontinue the use of the MAS feature and that the issue will be correct in the new Software Version 2.1 which is expected to be release in June 2020. -Fill out the Response Form and return by either: -fax (562) 240-6526 Attention: Technical Services -email to: technical [email protected] -mail to: ATTN: Technical Services, 11331 Valley View Street, Cypress, CA 90630 -Forward this communication to all required individuals with their organization. For further assistance with evaluating assay runs or for any questions, to contact Technical Services at +1-5672-240-6500, option 3 or via email [email protected]

Distribution

Worldwide distribution - U.S. Nationwide distribution including in the states of CO, FL, KS, MI, NC, PA, TX, and WA The countries of Austria, Czech Republic, France and Switerland.

Quantity

13943 plastic vials (24 vials per box)