FDA Recall Terminated

IntelliVue 1.4 GHz Remote Antenna Used with Philips MX4O Monitors-The affected product is a 74 foot (22.6m) combined coax and unshielded twisted pair data cable bundle and is a component of the Philips Remote Antenna (PN 867151). The Remote Antenna is used with IntelliVue MX40 Patient Worn Monitors in the USA, to extend the coverage area of a Core Access Point for the wireless Smart-hopping system.

Recall: Z-2547-2019 · Initiated August 23, 2019

Recall

Recall Number
Z-2547-2019
Event Number
83679
Firm
Philips North America, LLC
FEI Number
1218950
Product Code
MHX
Status
Terminated
Root Cause
Process control
Initiated
August 23, 2019
Terminated
January 3, 2024
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

IntelliVue 1.4 GHz Remote Antenna Used with Philips MX4O Monitors-The affected product is a 74 foot (22.6m) combined coax and unshielded twisted pair data cable bundle and is a component of the Philips Remote Antenna (PN 867151). The Remote Antenna is used with IntelliVue MX40 Patient Worn Monitors in the USA, to extend the coverage area of a Core Access Point for the wireless Smart-hopping system.

Reason

Remote Antenna Cable Used with Philips MX4O Monitors Does Not Have the Required Rating for Some Installations

Action

Philips issued customer notification on 08/19/2019 and delivering on 08/23/2019 via priority mail. Letter states reason for recall, health risk and action to take: Customers with affected cables installed need not take any immediate action. The cables subject to this recall do not affect the normal operation of MX4O System with Smart Hopping Infrastructure. Philips Healthcare will contact you to schedule a service team to replace affected cables. If you need any further information or support concerning this issue, please contact your local Philips representative or Philips Customer Support at 1-800- SUPPORT 722-9377.

Distribution

Nationwide

Quantity

1790 units