FDA Recall Terminated

Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K, MMT-723, MMT-723K, MMT-554, and MMT-754. MiniMed 530G: MMT-551, and MMT-751 Paradigm insulin infusion pumps are intended for use to deliver basal and bolus insulin to persons with diabetes.

Recall: Z-2544-2014 · Initiated March 13, 2014

Recall

Recall Number
Z-2544-2014
Event Number
68277
Firm
Medtronic MiniMed Inc.
FEI Number
3003166194
Product Code
OYC
Status
Terminated
Root Cause
Device Design
Initiated
March 13, 2014
Posted
August 22, 2014
Terminated
April 7, 2015
Address
18000 Devonshire St, Northridge, CA, 91325-1219

Description

Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K, MMT-723, MMT-723K, MMT-554, and MMT-754. MiniMed 530G: MMT-551, and MMT-751 Paradigm insulin infusion pumps are intended for use to deliver basal and bolus insulin to persons with diabetes.

Reason

Medtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion Pumps because they have received reports that users have accidentally programmed the pump to deliver the maximum bolus amount.

Action

The firm, Medtronic, sent an "IMPORTANT MEDICAL DEVICE SAFETY INFORMATION" letter dated March 13, 2014 to its customers. The letter described the product, problem and actions to be taken. The letter informs the customers that because accidental button pressing errors may occur it is important that customers always confirm the insulin dose flashing on the display is correct before pressing ACT to start delivery. The letter informs the customers of what safety settings will minimize the risk of an accidental programming error and what actions to be taken. Should you have any questions about this letter or concerns with your keypad, please do not hesitate to contact the 24 hour helpline at (866) 222-7304 for a replacement.

Distribution

Worldwide Distribution-Nationwide in US

Quantity

559,374 units total (444,374 units in US)