FDA Recall Open, Classified

RMU-2000 Automated Chest Compression Device

Recall: Z-2543-2024 · Initiated July 12, 2024

Recall

Recall Number
Z-2543-2024
Event Number
94960
Firm
Defibtech, LLC
FEI Number
3003521780
Product Code
DRM
Status
Open, Classified
Root Cause
Process control
Initiated
July 12, 2024
Posted
August 14, 2024
Address
741 Boston Post Rd, Suite 201, Guilford, CT, 06437-2714

Description

RMU-2000 Automated Chest Compression Device

Reason

Problem in the device's motor may cause the device to stop compressions. This may lead to a delay in therapy if the device fails to operate, and could lead to patient injury or death.

Action

On July 12, 2024, the firm notified affected users via Urgent Medical Device Safety Removal letters. Affected units (as identified by serial number) should be quarantined by the customer. Defibtech will contact customers to arrange for the return of the product. Customers' devices will be refunded, repaired, or replaced at no cost to them. Customers who have further transferred the device to a downstream account must notify those accounts of this notice or provide Defibtech with the contact information for those sub-accounts. Defibtech Customer Support: 1-877-453-4507, 7:30AM - 6PM Eastern, Monday thru Friday.

Distribution

Domestic distribution to CT, FL, NJ, SC, TN. Foreign distribution to Japan

Quantity

174 US, 37 OUS