FDA Recall Terminated

TRYX Absorbable Antibacterial Envelope Product Usage: Indicated for stabilization of implanted pacemakers (IPG) and/or implantable cardioverter defibrillators (ICD)

Recall: Z-2540-2016 · Initiated June 29, 2016

Recall

Recall Number
Z-2540-2016
Event Number
74672
Firm
TYRX Inc.
FEI Number
3005619263
Product Code
FTL
Status
Terminated
Root Cause
Process control
Initiated
June 29, 2016
Terminated
December 12, 2017
Address
1 Deerpark Dr, Ste G, Monmouth Junction, NJ, 08852-1920

Description

TRYX Absorbable Antibacterial Envelope Product Usage: Indicated for stabilization of implanted pacemakers (IPG) and/or implantable cardioverter defibrillators (ICD)

Reason

TRYX products are being recalled since the processes of spaying, welding, drying oven and polymer were not adequately validated.

Action

Medtronic sent an Urgent Medical Device Recall letter dated June 2016 to affected customers. The letter identified the affected product problem and actions to be taken. Customers are asked to immediately remove and quarantine all unused product that remains in inventory, return unused product to Medtronic and contacted the local Medtronic representative or Customer Service at 800-848-9300 to assist with the return and credit of unused product. Medtronic will provide credit for all non-expired, unused product. The customer should complete the attached Customer Confirmation Certificate and email to [email protected] or fax to Medtronic at 651-367-0612 Attention Customer Focused Quality.

Distribution

US Nationwide Distribution

Quantity

2993 units