FDA Recall Open, Classified

Berchtold Chromophare F528 Surgical Lights. Product Usage: The medical lights are for usage in hospital treatment rooms. They provide localized lighting focused on the patient s body for detecting and treating disease, injury, and disability. The surgical lights may only be used in rooms used for medical purposes that have been equipped in accordance with VDE 0100-710 or IEC 60364-7-10 in accordance with the regulations.

Recall: Z-2539-2019 · Initiated March 22, 2019

Recall

Recall Number
Z-2539-2019
Event Number
83355
Firm
Stryker Communications
FEI Number
3002975342
Product Code
FSY
Status
Open, Classified
Root Cause
Employee error
Initiated
March 22, 2019
Address
571 Silveron Blvd, Flower Mound, TX, 75028

Description

Berchtold Chromophare F528 Surgical Lights. Product Usage: The medical lights are for usage in hospital treatment rooms. They provide localized lighting focused on the patient s body for detecting and treating disease, injury, and disability. The surgical lights may only be used in rooms used for medical purposes that have been equipped in accordance with VDE 0100-710 or IEC 60364-7-10 in accordance with the regulations.

Reason

The back cover of the light head was insufficiently assembled and could potentially fall off the light.

Action

The recalling firm issued letters via FedEx on 3/22/2019 to their U.S. consignees. The recall letter identified the affected product, problem and actions to be taken. ,The letter instructs the customer to disseminate it within their organization and that a Stryker representative will begin replacements for customer facilities in January. For the replacement, the Stryker representative will perform an onsite rework of the light head cover. In the meantime, the lights can continue to remain in full use. A response form was enclosed for the customer to indicate they have read and understood the instructions. The form was to be returned via email. The recalling firm's representatives will contact the customer to schedule an inspection of their light and will track completion of the visits and inspections.

Distribution

Distribution was nationwide, including to Puerto Rico. There was also government/military distribution. Foreign distribution was made to Argentina, Austria, Australia, Belgium, Bulgaria, Canada, China, Colombia, Denmark, France, Germany, Hong Kong, India, Indonesia, Italy, Japan, Kazakhstan, Korea, Kuwait, Lithuania, Mexico, Netherlands, Philippines, Poland, Russia, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.

Quantity

946 lights