Stryker AutoPlex System. Model Numbers 0607-687-000, 0605-887-000, 0605-687-000, and 0605-683-000. For bone cement. Stryker AutoPlex System kits contain a cement mixer, cement injector, funnel, and an extension tube assembly whereby bone cement powder and liquid monomer are mixed inside a chamber and transferred into a cement injector automatically. The bone cement is then dispensed directly into a fracture site using the cement injector and introducer needle combination.
Recall
- Recall Number
- Z-2536-2016
- Event Number
- 74582
- Firm
- Stryker Instruments Div. of Stryker Corporation
- FEI Number
- 1811755
- Product Code
- NDN
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- June 14, 2016
- Posted
- August 15, 2016
- Terminated
- March 27, 2017
- Address
- 4100 E Milham Ave, Portage, MI, 49002-9704
Description
Stryker AutoPlex System. Model Numbers 0607-687-000, 0605-887-000, 0605-687-000, and 0605-683-000. For bone cement. Stryker AutoPlex System kits contain a cement mixer, cement injector, funnel, and an extension tube assembly whereby bone cement powder and liquid monomer are mixed inside a chamber and transferred into a cement injector automatically. The bone cement is then dispensed directly into a fracture site using the cement injector and introducer needle combination.
The Piston Head can become separated from the Delivery Piston, blocking the injection assembly Valve resulting in a cement backflow towards the injector handle. Potential for delay in surgery if additional cement needs to be prepared for the injection procedure.
On 6/13/2016, URGENT MEDICAL DEVICE RECALL NOTIFICATION notifications were sent to the affected customers via courier and sales representatives via email. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions regarding this recall, please contact Stryker Instruments: Jennifer Olson 269-389-2644 [email protected]
Worldwide distribution. US nationwide, France, Netherlands, Spain, Italy, Canada, Israel, Romania, Switzerland, and United Kingdom.
4,841 boxes (9,682 units)