FDA Recall Terminated

HexaPOD evo RT Couch Top (also known as HexaPOD evo Module), a component of the HexaPOD evo RT system. Product Usage: To support and aid in positioning a patient during radiation therapy

Recall: Z-2535-2018 · Initiated May 16, 2018

Recall

Recall Number
Z-2535-2018
Event Number
80471
Firm
Elekta, Inc.
FEI Number
1037831
Product Code
JAI
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
May 16, 2018
Terminated
October 30, 2020
Address
400 Perimeter Center Ter NE, Ste 50, Atlanta, GA, 30346-1227

Description

HexaPOD evo RT Couch Top (also known as HexaPOD evo Module), a component of the HexaPOD evo RT system. Product Usage: To support and aid in positioning a patient during radiation therapy

Reason

Upon completion of a 6D workflow, the HexaPOD evo Module may be in a tilted state if it is moved to the position used for 3D (non-iGUIDE) treatments resulting in incorrect patient positioning.

Action

The letter from Elekta dated May 2018 was issued via email 5/16/2018 and was flagged "URGENT IMPORTANT FIELD SAFETY NOTIFICATION. The letter identified the affected product, problem and actions to be taken.

Distribution

Worldwide Distribution - US Distribution of the device was made nationwide. There was government distribution but no military distribution. Foreign distribution was made to Canada, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Botswana, Brazil, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Democratic People's Republic of Korea, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Sweden, Thailand, Turkey, United Kingdom, and Viet Nam.

Quantity

725 devices