HexaPOD evo RT Couch Top (also known as HexaPOD evo Module), a component of the HexaPOD evo RT system. Product Usage: To support and aid in positioning a patient during radiation therapy
Recall
- Recall Number
- Z-2535-2018
- Event Number
- 80471
- Firm
- Elekta, Inc.
- FEI Number
- 1037831
- Product Code
- JAI
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- May 16, 2018
- Terminated
- October 30, 2020
- Address
- 400 Perimeter Center Ter NE, Ste 50, Atlanta, GA, 30346-1227
Description
HexaPOD evo RT Couch Top (also known as HexaPOD evo Module), a component of the HexaPOD evo RT system. Product Usage: To support and aid in positioning a patient during radiation therapy
Upon completion of a 6D workflow, the HexaPOD evo Module may be in a tilted state if it is moved to the position used for 3D (non-iGUIDE) treatments resulting in incorrect patient positioning.
The letter from Elekta dated May 2018 was issued via email 5/16/2018 and was flagged "URGENT IMPORTANT FIELD SAFETY NOTIFICATION. The letter identified the affected product, problem and actions to be taken.
Worldwide Distribution - US Distribution of the device was made nationwide. There was government distribution but no military distribution. Foreign distribution was made to Canada, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Botswana, Brazil, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Democratic People's Republic of Korea, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Sweden, Thailand, Turkey, United Kingdom, and Viet Nam.
725 devices